FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 12973321 · Received December 9, 2021

Report

Report Number
1024879-2021-00853
Event Type
Malfunction
Date Received
December 9, 2021
Date of Event
November 12, 2021
Report Date
January 28, 2022
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTOS WAS REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS NOT OBSERVED. THERE WERE NO DIFFICULTIES ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO DUPLICATE THE CUSTOMER¿S INDICATED FAILURE MODE, POOR BARRIER BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. ALL VISUAL OBSERVATIONS OF BOTH RETAIN AND CONTROL SAMPLES TESTED DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE. ALL TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION BASED ON PHOTOS ONLY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1280206, MEDICAL DEVICE EXPIRATION DATE: 2022-09-30, DEVICE MANUFACTURE DATE: 2021-10-07. MEDICAL DEVICE LOT #: 1253776, MEDICAL DEVICE EXPIRATION DATE. DEVICE MANUFACTURE DATE: 2021-09-10. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED POOR BARRIER SEPARATION OF SAMPLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT THE TUBES ARE NOT SEPARATING PROPERLY. EXPERIENCING ANOTHER ISSUE THIS TIME WITH GOLD TOPS. THEY ARE NOT COMPLETELY SEPARATING WHEN CENTRIFUGED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED POOR BARRIER SEPARATION OF SAMPLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT THE TUBES ARE NOT SEPARATING PROPERLY. EXPERIENCING ANOTHER ISSUE THIS TIME WITH GOLD TOPS. THEY ARE NOT COMPLETELY SEPARATING WHEN CENTRIFUGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1869467 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 367986 SEE H.10. 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Unknown