FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 12972446 · Received December 9, 2021

Report

Report Number
3013756811-2021-137956
Event Type
Malfunction
Date Received
December 9, 2021
Date of Event
November 24, 2021
Report Date
December 9, 2021
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613793
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. SYSTEM CHECK WAS PERFORMED BY TANDEM TECHNICAL SUPPORT AND NO ISSUES WERE IDENTIFIED. CUSTOMER SUCCESSFULLY RESUMED INSULIN DELIVERIES. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 190-457 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1871244 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1000354 00850006613793

Patients

Seq Age Sex Outcome Treatment
1 7 YR Male INSULIN: HUMALOG, INFUSION SET: AUTOSOFT 90