FDA Adverse Event Malfunction Summary report: N

UNI-VENT, MODEL 731

MDR report key: 12972109 · Received December 9, 2021

Report

Report Number
1220908-2021-04233
Event Type
Malfunction
Date Received
December 9, 2021
Report Date
November 18, 2021
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
CBK
UDI-DI
00847946023423
PMA / PMN Number
K111473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A FAULTY BOOTLOADER INTEGRATED CIRCUIT ON THE CENTRAL PROCESSING UNIT BOARD (CPU). THE THE CENTRAL PROCESSING UNIT BOARD WAS REPLACED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1871666 UNI-VENT, MODEL 731 VENTILATOR CBK ZOLL MEDICAL CORPORATION 799-EMVP-04-01 NA 00847946023423

Patients

Seq Age Sex Outcome Treatment
1 Unknown