FDA Adverse Event Injury Summary report: N

ENDOWRIST

MDR report key: 12965136 · Received December 8, 2021

Report

Report Number
2955842-2021-11698
Event Type
Injury
Date Received
December 8, 2021
Date of Event
November 16, 2021
Report Date
November 16, 2021
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874120767
PMA / PMN Number
K180351
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS REQUESTED THAT THE FORCE BIPOLAR INSTRUMENT BE RETURNED FOR FAILURE ANALYSIS TO BE PERFORMED, BUT THE INSTRUMENT HAS NOT YET BEEN RECEIVED. THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE HAS NOT BEEN DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT. NO IMAGE OR VIDEO CLIP FOR THE REPORTED EVENT WAS SUBMITTED FOR REVIEW. A REVIEW OF INSTRUMENT LOG WAS PERFORMED. THE INSTRUMENT WAS LAST USED ON (B)(6) 2021 ON SYSTEM (B)(4) FOR 14 MINUTES. FORCE BIPOLAR INSTRUMENT HAS 5 USES REMAINING AFTER LAST USE. THE INSTRUMENT HAS MAXIMUM 12 USES. THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, IT WAS ALLEGED THAT THE FORCE BIPOLAR INSTRUMENT PIN ON THE JAWS BECAME LOOSE AND FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE WITH NO PATIENT INJURY. AT THIS TIME IT IS UNKNOWN WHAT CAUSED THE BREAKAGE TO OCCUR. WHILE THERE WAS NO HARM OR INJURY TO THE PATIENT, THE REPORTED FAILURE MODE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF IT WERE TO RECUR AS UNINTENDED FRAGMENT(S) FALLING INTO THE PATIENT MAY REQUIRE SURGICAL INTERVENTION.

Additional Manufacturer Narrative · 0

BASED ON A RE-EVALUATION OF THE COMPLAINT INFORMATION, THIS COMPLAINT HAS BEEN RECLASSIFIED AS AN ADVERSE EVENT AND PRODUCT PROBLEM RATHER THAN JUST A PRODUCT PROBLEM, AS PREVIOUSLY REPORTED. CORRECTED INFORMATION CAN BE FOUND THE FOLLOWING FIELD: B1, B2, AND H1 B1 UPDATED FROM "PRODUCT PROBLEM" TO "ADVERSE EVENT AND PRODUCT PROBLEM". B2 UPDATED TO "REQUIRED INTERVENTION". H1 UPDATED FROM "MALFUNCTION" TO "SERIOUS INJURY".

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED VENTRAL HERNIA SURGICAL PROCEDURE, THE FORCE BIPOLAR INSTRUMENT GAGE PIN ON THE JAWS BECAME LOOSE AND THE HERNIA SAC WAS CAUGHT IN THE JAWS. BEFORE TAKING OUT THE INSTRUMENT, THE PIN FELL OUT INTO THE PATIENT. THE PIN WAS RETRIEVED SUCCESSFULLY WITH NO HARM TO THE PATIENT. THE INSTRUMENT WOULD NOT COME OUT TO THE CANNULA AND THE STAFF HAD TO ¿BURP¿ THE TROCAR OUT OF THE PATIENT. THE STAFF USED A SPARE INSTRUMENT AND PROCEEDED WITH THE CASE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE ISI CLINICAL SALES REPRESENTATIVE (CSR) AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: WHILE THE SURGEON WAS GETTING READY FOR HIS NEXT SURGICAL TASK, A PROTRUDING PIECE WAS NOTICED ON THE INSTRUMENT. THE SURGEON TOUCHED THE PIECE WITH ANOTHER INSTRUMENT AND IT FELL INSIDE OF THE PATIENT. IT IS UNKNOWN IF ANY TESTS OR X-RAYS WERE PERFORMED. THE SURGEON DOES NOT KNOW WHAT CAUSED THE INSTRUMENT TO BREAK. THE INSTRUMENT WAS INSPECTED PRIOR TO USE. THE INSTRUMENT DID NOT COLLIDE WITH ANY OTHER INSTRUMENTS OR HARD MATERIALS. THE INSTRUMENT WAS NOT REMOVED FROM THE PATIENT PRIOR TO THE BREAKAGE. THERE WAS NO HARM TO THE PATIENT. THE FRAGMENT WAS REMOVED DURING THE SAME PROCEDURE.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1856677 ENDOWRIST FORCE BIPOLAR NAY INTUITIVE SURGICAL, INC 471405-06 N10200914 0091 00886874120767

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.