FDA Adverse Event Malfunction Summary report: N

MOTOBAND CP IMPLANT SYSTEM

MDR report key: 12964836 · Received December 8, 2021

Report

Report Number
3011421599-2021-00014
Event Type
Malfunction
Date Received
December 8, 2021
Report Date
December 8, 2021
Manufacturer
CROSSROADS EXTREMITY SYSTEMS
Product Code
HRS
UDI-DI
00815432028870
PMA / PMN Number
K193452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

LOT HISTORY RECORDS (LHR'S) OF THE PLATE WAS REVIEWED. NON-CONFORMANCE WAS NOT IDENTIFIED BASED ON THE LOTS THAT WOULD CONTRIBUTE TO THIS FAILURE MODE. POTENTIALLY, FORCE FROM WEIGHT-BEARING WITH NORMAL WALKING MOTION CAUSED THE SCREW TO BACK-OUT. SCREW BACK-OUT IS A KNOWN FAILURE MODE OF SCREWS. REFER TO LOT NUMBER BELOW FOR ADDITIONAL COMPONENTS OF THE CONSTRUCT. REF: 7100-LP18-R, LN: 501122, QTY: 1, PRODUCT NAME: DYNABUNION RIGHT PLATE 18MM. REF: 7118-1814KT, LN: 300277, QTY: 1, PRODUCT NAME: 18X18X14MM HIMAX CLIP. REF: 15LP-3530, LN: 501163, QTY: 1, PRODUCT NAME: DYNABUNION ANTIDRIFT BOLT 30MM. REF: 15NL-3016, LN: 400118, QTY: 1, PRODUCT NAME: 3.0 X 16MM NON-LOCKING SCREW. REF: 1500-3514, LN: 400076, QTY: 1, PRODUCT NAME: 3.5 X 14MM NON-LOCKING SCREW. REF: 1500-3514, LN: 400007, QTY: 1, PRODUCT NAME: 3.5 X 14MM NON-LOCKING SCREW.

Description of Event or Problem · 0

THE SURGEON SUBMITTED A CASE SERIES OF DYNABUNION SYSTEM. THE ANTI-DRIFT BLOT HAS BACKED OUT AS SHOWN IN THE 6-WEEK X-RAY. THE SCREW HEAD PROMINENT ABOVE THE PLATE. THE PATIENT WAS ASYMPTOMATIC AND REVISION SURGERY WAS NOT SCHEDULED AT THE TIME OF THE REPORT. THE SURGEON REPORTED AT A LATER DATE THAT THE SYSTEM WAS REMOVED AFTER A PERIOD OF NON-WEIGHT BEARING. THE PATIENT STATUS AFTER THE REVISION IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1858009 MOTOBAND CP IMPLANT SYSTEM PLATE, FIXATION, BONE HRS CROSSROADS EXTREMITY SYSTEMS 7100-LP18-R 501122 00815432028870

Patients

Seq Age Sex Outcome Treatment
1 54 YR Female Required Intervention