FDA Adverse Event Malfunction Summary report: N

FREEDOM NEUROSTIMULATOR

MDR report key: 12963610 · Received December 8, 2021

Report

Report Number
3010676138-2021-00238
Event Type
Malfunction
Date Received
December 8, 2021
Date of Event
November 8, 2021
Report Date
December 7, 2021
Manufacturer
STIMWAVE TECHNOLOGIES INC.
Product Code
GZB
UDI-DI
00818225020150
PMA / PMN Number
K171366
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ALTHOUGH THE PEELING IS VISIBLE, THE REPORTED ISSUE OF OVERHEATING AND THE CAUSE OF THE PEELING COULD NOT BE REPLICATED. FOR THE REPORT OF PEELING: VISUAL INSPECTION OF UNIT SHOWS UNUSUAL PATTERN OF EXTERNAL CORROSION, INCLUDING PAINT PEELING OFF. THE CORROSION LIKE EFFECT ALSO PROPAGATED INTO THE INSIDE OF THE ENCLOSURE. PER THE INSTRUCTIONS FOR USE FOR THE WEARABLE ANTENNA ASSEMBLY, THE APPROVED CLEANING METHOD IS THE FOLLOWING: CLEAN WITH A DAMP CLOTH, MILD HOUSEHOLD CLEANERS WILL NOT DAMAGE THE DEVICE OR LABELS. THIS HAS BEEN VALIDATED BY STIMWAVE PER (B)(4) -CLEANING AND DURABILITY OF MARKINGS. BASED ON THIS TEST REPORT AND INSTRUCTIONS FOR USE, THE WAA HAS BEEN TESTED TO WITHSTAND MILD HOUSEHOLD CLEANERS (SUCH AS ISOPROPYL ALCOHOL, BLEACH , SOAP, ETC). THIS IS THE FIRST REPORTED COMPLAINT FOR PEELING SINCE COMMERCIAL DISTRIBUTION. FOR THE REPORT OF OVERHEATING: THE PCB IS IN GOOD CONDITION, WITH THE EXCEPTION OF THE BROKEN RF CONNECTOR SOLDER JOINT. THERE IS NO EVIDENCE OF OVERHEATING ON THE PCB. THE OUTSIDE THERMAL TEMPERATURE WHILE CHARGING WAS 29.0°C, WHICH IS BELOW THE PRODUCT REQUIREMENTS SPECIFICATION OF 41°C (106°F). THE PA FAILED OPEN WHICH DOES NOT CAUSE EXCESSIVE CURRENT DRAW. THE DEVICE CANNOT OVERHEAT BECAUSE THERE IS NO CURRENT DRAW. THE LIKELY CAUSE OF PA FAILURE IS THE BROKEN RF CONNECTOR. THERE IS EVIDENCE OF INTRUSION OF FOREIGN MATTER INTO THE ENCLOSURE, AN UNIDENTIFIED PARTICULATE IN THE VICINITY OF THE USB CONNECTOR AND THE RF CONNECTOR. THE UNIDENTIFIED PARTICULATE DID NOT CAUSE THE REPORTED ISSUE. ALTHOUGH THE REPORTED ISSUE COULD NOT BE REPLICATED, THE INVESTIGATION FOUND THE RF CONNECTOR WAS DAMAGED. BASED ON THIS INFORMATION, THE COMPLAINT WILL BE CLOSED AS NO PROBLEM FOUND FOR THE REPORTED ISSUES OF PEELING AND OVERHEATING.

Description of Event or Problem · 0

THE PATIENT REPORTED THEIR WEARABLE ANTENNA ASSEMBLY GETS HOT AND THE OUTSIDE IS PEELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1856220 FREEDOM NEUROSTIMULATOR SPINAL CORD NERVE STIMULATOR GZB STIMWAVE TECHNOLOGIES INC. PDBT-915-2K 00365-D 00818225020150

Patients

Seq Age Sex Outcome Treatment
1 Unknown