FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE

MDR report key: 12962583 · Received December 8, 2021

Report

Report Number
1911916-2021-01232
Event Type
Malfunction
Date Received
December 8, 2021
Date of Event
September 27, 2021
Report Date
January 17, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 2022-01-31. INVESTIGATION SUMMARY: TO AID IN THE INVESTIGATION, ONE SAMPLE IN AN OPENED PACKAGING BLISTER AND ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THE RETURNED SAMPLE HAS THE SCALE MARKING SKEWED. THE PHOTO SHOWS A SYRINGE WITH NO PACKAGING BLISTER. THE SYRINGE BARREL HAS THE SCALE MARKING SKEWED. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THERE WAS A JAM DURING THE PRINTING PROCESS INDUCING THE SYMPTOM REPORTED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 303552, LOT NUMBER 1103477. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. VERIFICATION OF THE PRINTING PROCESS WAS PERFORMED. THE SETTINGS WERE CORRECT AND THE FLOW OF PRODUCT WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ SYRINGE SCALE MARKINGS WERE CROOKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "CROOKED GRADUATION.".

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ SYRINGE SCALE MARKINGS WERE CROOKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "CROOKED GRADUATION."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ SYRINGE SCALE MARKINGS WERE CROOKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "CROOKED GRADUATION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1858850 BD LUER-LOK¿ SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 1103477

Patients

Seq Age Sex Outcome Treatment
1 Unknown