ADVIA CALCIUM_2 ASSAY
Report
- Report Number
- 2432235-2009-00013
- Event Type
- Other
- Date Received
- January 23, 2009
- Date of Event
- January 7, 2009
- Report Date
- January 7, 2009
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K974607
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CUSTOMER HAD CHANGED THE REAGENT WEDGE DURING NORMAL TROUBLESHOOTING PROCEDURES, THE SYSTEM WAS CALIBRATED AND QC WAS RUN WITH ALL RESULTS BEING NORMAL. A SIEMENS HEALTHCARE DIAGNOSTIC INC. FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT. HE CHECKED THE SYSTEM OUT, AND RAN SEVERAL QC TESTS. ALL CAME OUT NORMAL. THE DEVICE IS OPERATING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
CUSTOMER REPORTED INTERMITTENTLY LOW ADVIA 1650 CALCIUM CA_2 PT RESULTS TO THE PHYSICIAN. THE QC RUN DATA WAS ACCEPTABLE. CUSTOMER REPEATED THE SAMPLE ON AN ALTERNATE ANALYZER AND THE CALCIUM RESULTS WERE HIGHER. PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT CALCIUM RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVIA CALCIUM_2 ASSAY | CALCIUM ARSENAZO III METHOD | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |