FDA Adverse Event Malfunction Summary report: N

ATUNE REV RP TIB BASE SZ 4 CEM

MDR report key: 12954980 · Received December 7, 2021

Report

Report Number
1818910-2021-27419
Event Type
Malfunction
Date Received
December 7, 2021
Date of Event
January 1, 2021
Report Date
December 7, 2021
Manufacturer
DEPUY IRELAND - 9616671
Product Code
NJL
UDI-DI
10603295042884
PMA / PMN Number
P830055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # ==> (B)(4) INVESTIGATION SUMMARY ==> NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT ==> A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE 150660004, LOT 9915805, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED. JE (B)(6) 2022

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOWUP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SURGEON WAS IMPACTING THE REVISION TIBIAL COMPONENTS IN THE REVISION TIBIA CONSTRUCT SAT UP SEVERAL MM FROM WHERE THE TRIAL SAT WHEN CEMENTING. HE HAD TO REMOVE TIBIA CLEAR OUT ALL THE CEMENT THAT HE JUST MIXED FOR IMPLANTS BECAUSE IT HARDENED. WE HAD TO OPEN NEW TIBIAL COMPONENTS FOR PATIENT WHICH CAUSED A 35-40 MIN DELAY IN THE CASE. DOI: (B)(6) 2021, DOE: (B)(6) 2021, RIGHT KNEE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1853762 ATUNE REV RP TIB BASE SZ 4 CEM KNEE TIBIAL TRAY NJL DEPUY IRELAND - 9616671 1506-60-004 9915805 10603295042884

Patients

Seq Age Sex Outcome Treatment
1 Unknown ATTUNE CEMENTED STEM 14X50MM| ATUN TIB SLV M/L 29MM HALF POR| ATUNE REV RP TIB BASE SZ 4 CEM