FDA Adverse Event Malfunction Summary report: N

X-SMART PLUS

MDR report key: 12954845 · Received December 7, 2021

Report

Report Number
8031010-2021-00376
Event Type
Malfunction
Date Received
December 7, 2021
Date of Event
November 22, 2021
Report Date
March 30, 2022
Manufacturer
MAILLEFER INSTRUMENTS HOLDING SARL
Product Code
EGS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE UNSUCCESSFUL ATTEMPTS WERE MADE TO OBTAIN THE DEVICE FOR EVALUATION.

Additional Manufacturer Narrative · 0

DEVICE FOR THIS COMPLAINT WAS RECEIVED AND THE INVESTIGATION/INSPECTION RESULTS FOR THIS COMPLAINT ARE RECEIVED = 1X X-SMART PLUS BEARING RETAINER H1033C1051113 1X X-SMART PLUS CARTRIDGE H1033C1051015 1X X-SMART PLUS DRIVE SHAFT H1033C1051022 1X X-SMART PLUS HEAD CAP H1033C1051050 1X X-SMART PLUS WAVE WASHER H1033C427A360 X-SMART PLUS CARTRIDGE BAD MAINTENANCE (USER) THERE IS SOME RUST ON THE CARTRIDGE.

Additional Manufacturer Narrative · 0

CONTRA ANGLE WON'T HOLD FILES; NO INJURY RESULTED. THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT X-SMART PLUS WAS NOT HOLDING FILES AND HAD BAD SOUND. THERE IS NO INJURY TO PATIENT. AS PER CUSTOMER: SOUND / FILE NOT HOLDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1843986 X-SMART PLUS HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL EGS MAILLEFER INSTRUMENTS HOLDING SARL 8689368

Patients

Seq Age Sex Outcome Treatment
1 Unknown