REMSTAR AUTO A-FLEX
Report
- Report Number
- 2518422-2021-07808
- Event Type
- Malfunction
- Date Received
- December 7, 2021
- Date of Event
- November 8, 2021
- Report Date
- June 26, 2023
- Manufacturer
- RESPIRONICS,INC
- Product Code
- BZD
- UDI-DI
- 00606959005884
- PMA / PMN Number
- K131982
- Removal / Correction Number
- RES 88058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION OF AN ISSUE RELATED TO SOUND ABATEMENT FOAM. A CORRECTION TO B5 WAS MADE AND SHOULD BE: THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION VISUALIZATION OF PARTICLES RELATED TO CPAP DEVICE'S SOUND ABATEMENT FOAM AND BECAME DEGRADED AND CAUSED THE PATIENT TO HAVE BLOODY NOSE AND DRY MOUTH. THE MEDICAL INTERVENTION THAT THE PATIENT RECEIVED IN RESPONSE TO THE EVENT IS CURRENTLY UNKNOWN. REPEATED ATTEMPTS TO HAVE THE DEVICE AND COMPONENTS RETURNED FOR EVALUATION AND INVESTIGATION WERE UNSUCCESSFUL. THE MANUFACTURER BELIEVES THEY WILL BE UNABLE TO GATHER ADDITIONAL INFORMATION. THE MANUFACTURER IS SUBMITTING AN UPDATED REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, A FOLLOW-UP REPORT WILL BE FILED. SECTION H6 UPDATED IN THIS REPORT.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1849795 | REMSTAR AUTO A-FLEX | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS,INC | DS560S | 00606959005884 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |