FDA Adverse Event Malfunction Summary report: N

CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR

MDR report key: 1293483 · Received January 26, 2009

Report

Report Number
2018433-2009-00007
Event Type
Malfunction
Date Received
January 26, 2009
Date of Event
October 23, 2008
Report Date
October 23, 2008
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JIX
PMA / PMN Number
K981706
Removal / Correction Number
2018433-09/19/08-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: ANALYZER. EVALUATION RESULT: EXPLANATION: ABBOTT CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR VALUE ASSIGNMENT PROCESS WAS INADEQUATE CAUSING LOW AND HIGH RESULTS ON PATIENT SAMPLES AND CAP SURVEYS. ABBOTT CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR IS USED ON THE AEROSET AND ARCHITECT CSYSTEM TO CALIBRATE TOTAL AND DIRECT BILIRUBIN. ABBOTT CUSTOMERS REPORTED THE FOLLOWING ISSUES FOR CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR: - LOW BIAS FOR TOTAL BILIRUBIN QUALITY CONTROL (QC) AND PATIENT RESULTS WITH BILIRUBIN CALIBRATOR LOT NUMBER CHANGE - RESULTS TRENDING LOW WITH PROFICIENCY SURVEY SAMPLES. INVESTIGATION OF THESE ISSUES DETERMINED THAT THE CALIBRATOR VALUES ASSIGNED TO THE TOTAL BILIRUBIN REAGENT CAUSED RESULTS TO TREND DOWNWARD OVER THE PAST SEVERAL CALIBRATOR LOT NUMBERS. - WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 41456M100 A BIAS OF (-) 7% TO 3% WAS OBSERVED. - WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 42396M100 A BIAS OF (-) 4% TO 2% WAS OBSERVED. - WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 48616M100 A BIAS OF (-) 5.5% TO 3.2% WAS OBSERVED. ON AUGUST 10, 2007, A PRODUCT CORRECTION LETTER WAS ISSUED TO ADDRESS BILIRUBIN LOW RECOVERY COMPLAINTS AND PROVIDED CUSTOMERS WITH NEW BILIRUBIN VALUES FOR ALL ON-MARKET CALIBRATOR LOTS. THE NEW BILIRUBIN VALUES WERE ASSIGNED USING THE SECONDARY STANDARD (VERICHEM). NIST STANDARD PREPARATION WAS IDENTIFIED AS A SOURCE OF VARIABILITY. THE VERICHEM STANDARD MINIMIZED LOT-TO-LOT VARIABILITY IN THE VALUE ASSIGNMENT PROCESS. FOLLOWING THE USE OF VERICHEM AS A STANDARD FOR THE ASSIGNMENT FOR THE BILIRUBIN CALIBRATOR, AN INCREASE IN CUSTOMER COMPLAINTS FOR HIGH RESULTS FOR TOTAL BILIRUBIN REAGENTS WAS OBSERVED: - HIGHER THAN EXPECTED RESULTS ON PROFICIENCY SURVEY SAMPLES - HIGHER THAN EXPECTED QUALITY CONTROL (QC) RECOVERY - HIGHER THAN EXPECTED PATIENT RESULTS AN INVESTIGATION WAS INITIATED FOR THE TOTAL BILIRUBIN REAGENTS WHICH DETERMINED THAT THE MATRIX OF THE SECONDARY STANDARD USED IN THE VALUE ASSIGNMENT OF THE CALIBRATOR IS SENSITIVE TO THE DIAZO METHOD IN THE ABBOTT TOTAL BILIRUBIN REAGENTS. THE MATRIX OF THIS SECONDARY STANDARD CAUSED A POSITIVE BIAS. BECAUSE OF THESE FINDINGS, A NEW TOTAL BILIRUBIN CALIBRATOR VALUE ASSIGNMENT PROCESS WAS IMPLEMENTED FOR USE WITH THE TOTAL BILIRUBIN REAGENTS. THE REVISED VALUES LOWERED TOTAL BILIRUBIN RESULTS UP TO 18%. THERE IS NO ESTABLISHED REFERENCE METHOD FOR TOTAL BILIRUBIN, HOWEVER LITERATURE REFERENCES DOCUMENT ONGOING EFFORTS OF STANDARDIZATION. ABBOTT HAS SELECTED A TOTAL BILIRUBIN CALIBRATOR VALUE ASSIGNMENT PROCESS USED BY THE COLLEGE OF AMERICAN PATHOLOGISTS (CAP) FOR NEONATAL PROFICIENCY SURVEY SAMPLES. ACCURACY OF THE NEW PROCESS WAS CONFIRMED USING THE JENDRASSIC-GROFF METHOD MODIFIED BY DOUMAS. THE IMPROVED VALUE ASSIGNMENT PROCEDURE REMAINS TRACEABLE TO NIST AND DEMONSTRATES MINIMAL SENSITIVITY TO MATRIX VARIABILITY. ALIGNMENT OF THE STANDARDIZATION DIRECTLY WITH CONFIRMATION OF TOTAL BILIRUBIN COMMUTABILITY TO THE DOUMAS JENDRASSIC-GROFF METHOD IMPROVES THE VALUE ASSIGNMENT ACCURACY. THE ROOT CAUSE WAS DETERMINED TO BE MATRIX ISSUES FOR TOTAL BILIRUBIN REAGENT (DIAZO) RELATED TO THE COMMUTABILITY OF VARIOUS SAMPLES (NIST STANDARD, VERICHEM STANDARD, BILIRUBIN CALIBRATOR, AND PATIENT SAMPLES) DUE TO THE PRODUCTION MATERIALS USED, E.G. HUMAN SERUM ALBUMIN (HSA) VS. BOVINE SERUM ALBUMIN (BSA). THIS MATRIX EFFECT WAS RESPONSIBLE FOR THE OVER-RECOVERY OBSERVED BY CUSTOMERS DURING THE CONVERSION TO THE VERICHEM STANDARD. THE CAUSE OF THIS ISSUE WAS CONFIRMED THROUGH TESTING PERFORMED USING AN EXTERNAL REFERENCE JENDRASSIC AND GROFF METHOD. AS A CORRECTIVE/PREVENTIVE MEASURE ALL ON-MARKET BILIRUBIN CALIBRATOR LOTS WERE RE-VALUE ASSIGNED PER THE NEW VALUE ASSIGNMENT PROCEDURE PROVIDING IMPROVED ACCURACY FOR TOTAL BILIRUBIN REAGENTS. THESE VALUES WERE COMMUNICATED IN THE CUSTOMER LETTER THAT WAS ISSUED ON SEPTEMBER 17, 2008 (FA12SEP2008 REVISION 01). IN ADDITION, ABBOTT HAS RE-WRITTEN AND VALIDATED THE PROCEDURE FOR BILIRUBIN CALIBRATOR. VALUE ASSIGNMENT AND VALUE VERIFICATION THAT INCLUDE: - CLARIFIED PROCEDURE FOR PRIMARY NIST STANDARD PREPARATION AND INCREASING REPLICATES TO IMPROVING ACCURACY; REVALIDATED THE PROCEDURE. - REASSIGNED OF VERICHEM STANDARDS USING IN-HOUSE PREPARED NIST STANDARD TO ALLOW FOR LESS VARIABILITY FROM ONE ASSIGNMENT TO ANOTHER. - ASSIGNED THE VALUES FOR BILIRUBIN CALIBRATOR USING REASSIGNED VERICHEM STANDARDS BY IN-HOUSE PREPARED NIST STANDARDS PER JENDRASSIC AND GROFF METHOD. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

THE CUSTOMER STATED THEIR DISAGREEMENT WITH THE NEW CONCENTRATION VALUE OF THE TOTAL BILIRUBIN CALIBRATORS IN THE CUSTOMER COMMUNICATION LETTER ASSOCIATED WITH FIELD ACTION FA12SEP2008. THE CUSTOMER STATED THEIR BIORAD CONTROL LEVEL 1 DROPPED FROM 0.916 TO 0.74 MG/DL, AND THE LEVEL 2 DROPPED FROM 4.83 TO 3.9 MG/DL. WITH THE NEW CALIBRATION VALUES, THEY WOULD BE REPORTING LOWER VALUES WITH A SIGNIFICANT BIAS. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR FOR USE IN THE CALIBRATION OF THE CLINICAL CHEMISTRY BILIRUBIN ASSAY JIX ABBOTT MANUFACTURING, INC. 57919M100

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT C16000 ANALYZER