FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 12933002 · Received December 6, 2021

Report

Report Number
1000113657-2021-00759
Event Type
Malfunction
Date Received
December 6, 2021
Date of Event
November 9, 2021
Report Date
January 5, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED SHE HAD NOT YET RECEIVED THE REPLACEMENT PRODUCTS. CUSTOMER STATED THEY WOULD CALL IF FURTHER ASSISTANCE IS NEEDED.

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 05-JAN-2022. H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS. H10: METER WAS NOT RETURNED FOR EVALUATION. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN 6 MONTHS TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-055: USER HAD AN INACCURATE REFERENCE: COMPETITOR¿S METER: THE END USER IS COMPARING RESULTS OBTAINED FROM TRIVIDIA¿S BGM SYSTEM TO THE RESULTS FROM A COMPETITOR¿S BGM SYSTEM.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. SON IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 445, 421, 345 AND 459 MG/DL. THE CUSTOMER IS ALSO CONCERNED WITH FASTING METER TO METER COMPARISON RESULTS OF 445 MG/DL USING TRUE METRIX METER AND 160 MG/DL USING ANOTHER DEVICE. THE CUSTOMER DOES NOT KNOW HER EXPECTED BLOOD GLUCOSE TEST RESULT RANGE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 419 MG/DL USING TRUE METRIX METER; THE CUSTOMER WAS NOT SATISFIED WITH THE RESULT OBTAINED. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 09/30/2022 AND OPEN VIAL DATE IS (B)(6) 2021. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER ONLY HAS FOUR RESULTS IN MEMORY): (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1829974 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX HNO 50CT24/CS MG/DL ZY4306S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown