FDA Adverse Event Injury Summary report: N

ALPINE LEG BAG LARGE 32OZ

MDR report key: 1293102 · Received January 21, 2009

Report

Report Number
2183558-2009-00006
Event Type
Injury
Date Received
January 21, 2009
Date of Event
October 1, 2008
Report Date
October 29, 2008
Manufacturer
COLOPLAST MANUFACTURING US LLC
Product Code
FAQ
PMA / PMN Number
K934052
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THREE LEG BAGS WERE RECEIVED FOR EVAL. EXAMINATION AND TESTING OF THE RETURNED PRODUCTS CONFIRMED THE COMPLAINT AS REPORTED WITH ONE OF THE LEG BAGS. WHEN WATER WAS INTRODUCED INTO THE THREE BAGS, WATER FLOWED FREELY INTO TWO BAGS, BUT DID NOT FLOW INTO ONE BAG AS EXPECTED. THE BAG WAS FURTHER EXAMINED AND IT WAS NOTED THAT THE ANTI-REFLUX VALVE WAS NOT WORKING PROPERLY. PERSONNEL TRIED OPENING THE VALVE AND DID SO MANUALLY. THE MALFUNCTIONING BAG WAS THEN SENT TO THE MFR FOR FURTHER EVAL. RESULTS FROM THAT EXAMINATION INDICATED THAT THE MFR WAS NOT ABLE TO DUPLICATE THE MALFUNCTION, AS THE BAG HAD BEEN MANUALLY OPENED PRIOR TO THE MFR RECEIVING IT. HOWEVER, IT WAS NOTED THAT THE SLIT APPEARED CENTERED AND COMPLETE. THE CONTRACT MFR IS GOING TO PUT ADDITIONAL TESTING IN PLACE AT INCOMING INSPECTION TO VERIFY THE VALVE IS WORKING CORRECTLY. TO DATE, THERE HAVE BEEN NO OTHER COMPLAINTS RECORDED FOR THIS LOT NUMBER.

Description of Event or Problem · 1

ACCORDING TO THE INFO RECEIVED, AN END-USER REPORTED THAT THE LAST 4 OF THE 68001'S WERE DEFECTIVE, AS URINE WOULD NOT DRAIN IN THEM. THE URINE WOULD COLLECT ON TOP OF THE ANTI-REFLUX VALVE (3 OF THE DEFECTIVE PRODUCTS WERE LOT 08119 AND 1 WAS LOT 08182) SO IT MAY BE DEFECTIVE VALVES. THE PT'S WIFE TOOK HIM TO ER BECAUSE OF CONCERN THAT THE FOLEY WAS NOT IN OR NOT WORKING PROPERLY. ONCE AT THE ER, HIS BLADDER WAS DRAINED AND NO KNOWN INJURY WAS REPORTED. PT WENT HOME FROM HOSP AND STILL HAD DRAINAGE ISSUES WITH THE BAGS, SO CALLED COLOPLAST TO SEE IF THEY WERE CONNECTING THE BAG CORRECTLY TO THE FOLEY. WHILE ON THE PHONE, THE WIFE SAID THAT SHE COULD SEE THE URINE IN THE TUBING BUT ONCE SHE DISCONNECTED THE TUBING FROM THE BAG, THE URINE WAS SITTING ON TOP OF THE ANTI REFLUX VALVE AND THEN THERE WAS NORMAL URINE FLOW DRAINING FROM THE FOLEY WHEN DETACHED FROM LEG BAG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALPINE LEG BAG LARGE 32OZ LEG BAG FAQ COLOPLAST MANUFACTURING US LLC 5596001400 08119

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention