FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1712KL 640G V4.10 BK SF MG

MDR report key: 12930437 · Received December 4, 2021

Report

Report Number
2032227-2021-226640
Event Type
Malfunction
Date Received
December 4, 2021
Date of Event
November 25, 2021
Report Date
July 6, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
000000763000317195
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

RETAINER RING=BLACK. CUSTOMER COMPLAINED ON 11/25/2021 PUMP ALARMED CRITICAL PUMP ERROR AFTER MOISTURE EXPOSURE. DEVICE ALARMED CRITICAL PUMP ERROR AFTER BATTERY INSTALLATION. PUMP ERROR 35 WAS NOTED IN THE PUMP HISTORY DOWNLOAD ON 11/25/2021 17:08:34.000 DUE TO MOISTURE DAMAGE TO FORCE SENSOR. DEVICE SUCCESSFULLY DOWNLOADED TO THUS AND CREST. UNABLE TO PERFORM SELFTEST, REWIND, SEATING, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST, ACTIVE CURRENT MEASUREMENT AND SLEEP CURRENT MEASUREMENT TEST DUE TO CRITICAL PUMP ERROR. FOUND MOISTURE DAMAGE TO MOTOR ASSEMBLY, PCB1 BOARD AND PCB2 BOARD DURING VISUAL INSPECTION. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED ELECTRONIC ASSEMBLIES, CORRODED BATTERY TUBE, CORRODED MOTOR HOME SWITCH, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED BATTERY TUBE THREADS AND CRACKED KEYPAD OVERLAY. THE P-CAP/RESERVOIR DOES LOCK PROPERLY. CONFIRMED CRITICAL PUMP ERROR AFTER BATTERY INSTALLATION AND PUMP ERROR 35 IN THE PUMP HISTORY DOWNLOAD ON 11/25/2021 17:08:34.000 DUE TO MOISTURE DAMAGE TO FORCE SENSOR. CONFIRMED MOISTURE DAMAGE TO MOTOR ASSEMBLY, PCB1 BOARD AND PCB2 BOARD DURING VISUAL INSPECTION. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR FURTHER ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1828934 PUMP MMT-1712KL 640G V4.10 BK SF MG PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1712KL HG4GT96 000000763000317195

Patients

Seq Age Sex Outcome Treatment
1 Unknown