FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 12923939 · Received December 3, 2021

Report

Report Number
1000113657-2021-00757
Event Type
Malfunction
Date Received
December 3, 2021
Date of Event
November 8, 2021
Report Date
December 3, 2021
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). TEST STRIPS WERE NOT RETURNED FOR EVALUATION. METER WAS RETURNED. REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058:USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR LOW AND HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 152, 336, 142, 294 AND 261MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 130-150MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 81MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 08/31/2022 AND OPEN VIAL DATE IS (B)(6) 2021. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 152MG/DL, DATE: (B)(6) 2021, TIME: 3:39PM, FASTING. RESULT 2: 336MG/DL, DATE: (B)(6) 2021, TIME: 2:58AM, NON-FASTING. RESULT 3: 142MG/DL, DATE: (B)(6) 2021, TIME: 2:59PM, NON-FASTING. RESULT 4: 294MG/DL, DATE: (B)(6) 2021, TIME: 3:24PM, UNKNOWN. RESULT 5: 261MG/DL, DATE: (B)(6) 2021, TIME: 9:03PM, UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1819878 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL ZX4277S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown