FDA Adverse Event Malfunction Summary report: N

GAP SPACER

MDR report key: 12919510 · Received December 2, 2021

Report

Report Number
1020279-2021-08488
Event Type
Malfunction
Date Received
December 2, 2021
Date of Event
November 15, 2021
Report Date
February 18, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWT
UDI-DI
00885556578445
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE CASE-(B)(4).

Additional Manufacturer Narrative · 0

H3, H6: THE DEVICE WAS NOT RETURNED FOR EVALUATION AND THE REPORTED EVENT COULD NOT BE CONFIRMED. A REVIEW OF COMPLAINT HISTORY REVEALED SIMILAR EVENTS FOR THE LISTED DEVICE, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. A HISTORICAL REVIEW CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS PRODUCT AND EVENT. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE CORROBORATED. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. WITHOUT THE RETURN OF THE ACTUAL PRODUCT INVOLVED, OUR INVESTIGATION COULD NOT PROCEED. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY.

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING A TKA SURGERY, A GAP SPACER BROKE. NO PIECES FELL INTO PATIENT. SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1817345 GAP SPACER TEMPLATE HWT SMITH & NEPHEW, INC. 71935180 UNK 00885556578445

Patients

Seq Age Sex Outcome Treatment
1 Unknown