FDA Adverse Event Malfunction Summary report: N

MICRO CHPV W RICKHAM UNITIZED

MDR report key: 12917616 · Received December 2, 2021

Report

Report Number
3013886523-2021-00526
Event Type
Malfunction
Date Received
December 2, 2021
Date of Event
November 8, 2021
Report Date
April 8, 2022
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
UDI-DI
10886704041481
PMA / PMN Number
K053107
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: D4, D9, G3, G6, H2, H3, H6, H10. HAKIM VALVE (823113) WAS NOT RETURNED FOR EVALUATION AFTER THREE GOOD FAITH ATTEMPTS (GFES) WERE MADE. LOT NUMBER INFORMATION HAS NOT BEEN PROVIDED; THEREFORE, AN EVALUATION OF THE DEVICE COULD NOT BE PERFORMED, AND DEVICE HISTORY RECORD (DHR) COULD NOT BE REVIEWED. THE ROOT CAUSE(S) OF THE REPORTED ISSUE COULD NOT BE DETERMINED. HOWEVER, A PROBABLE ROOT CAUSE FOR THE REPORTED COMPLAINT COULD BE DUE TO BIOLOGICAL DEBRIS AND PROTEIN BUILD UP INTERFERING WITH THE VALVE MECHANISM. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.

Description of Event or Problem · 0

A FACILITY REPORTED A HAKIM VALVE BECAME STUCK OVER TIME. IT IS UNKNOWN WHETHER THIS IS AN ISSUE CAUSED BY THE VALVE ITSELF, OR BY OTHER FACTORS. THE PATIENT MAY NEED REVISION SURGERY TO REPLACE THE VALVE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1815770 MICRO CHPV W RICKHAM UNITIZED CHPV JXG INTEGRA LIFESCIENCES SWITZERLAND SAR 82-3113 10886704041481

Patients

Seq Age Sex Outcome Treatment
1 Unknown