FDA Adverse Event Injury Summary report: N

TENDRIL STS

MDR report key: 12913804 · Received December 1, 2021

Report

Report Number
3008377825-2021-00020
Event Type
Injury
Date Received
December 1, 2021
Date of Event
November 17, 2021
Report Date
April 20, 2022
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Product Code
NVN
UDI-DI
05414734502870
PMA / PMN Number
P960013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: D4 : CHANGED THE SERIAL NUMBER OF THE PRODUCT FROM (B)(6) TO (B)(6). ADDITIONAL INFORMATION: THE REPORTED EVENTS WERE LEAD DISLODGEMENT AND HIGH THRESHOLD. AS RECEIVED, A COMPLETE LEAD WAS RETURNED IN ONE PIECE WITH THE HELIX RETRACTED AND CLOGGED WITH BLOOD/TISSUE. AFTER CLEANING, THE HELIX COULD BE EXTENDED AND RETRACTED WHEN TORQUE APPLIED TO THE CONNECTOR PIN. THE FULL HELIX EXTENSION LENGTH WAS MEASURED WITHIN SPECIFICATIONS. THE REPORTED EVENT OF HIGH THRESHOLD WAS NOT CONFIRMED. ELECTRICAL TESTING DID NOT FIND ANY INDICATION OF CONDUCTOR FRACTURES OR INTERNAL SHORTS. VISUAL AND X-RAY EXAMINATION DID NOT FIND ANY ANOMALIES WITH THE EXCEPTION OF PROCEDURAL DAMAGE.

Additional Manufacturer Narrative · 0

CORRECTION: D3 - CHANGED MANUFACTURER INFORMATION ADDRESS TO ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FROM ST. JUDE MEDICAL OPERATIONS (M) SDN BHD. ADDITIONAL INFORMATION - H6- CONCLUSION CODES.

Description of Event or Problem · 0

IT WAS REPORTED THE PATIENT PRESENTED IN CLINIC WITH DISCOMFORT FROM PACING. UPON INTERROGATION, IT WAS OBSERVED THE PATIENT'S ATRIAL LEAD HAD HIGH CAPTURE THRESHOLDS. A CHEST X-RAY SHOWED THE PATIENT'S ATRIAL LEAD HAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED WITHOUT ISSUE. THE PATIENT WAS STABLE THROUGHOUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1804891 TENDRIL STS PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC.(CRM-SYLMAR) 2088TC/46 P000117260 05414734502870

Patients

Seq Age Sex Outcome Treatment
1 78 YR Female Required Intervention ASSURITY PACEMAKER| TENDRIL STS LEAD