FDA Adverse Event Malfunction Summary report: N

MRIDIAN LINAC

MDR report key: 12913779 · Received December 1, 2021

Report

Report Number
3011233554-2021-00002
Event Type
Malfunction
Date Received
December 1, 2021
Date of Event
November 21, 2019
Report Date
December 1, 2021
Manufacturer
VIEWRAY, INC.
Product Code
IYE
UDI-DI
10868029000305
PMA / PMN Number
K170751
Removal / Correction Number
30112335544-11-18-19-003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS A RETROSPECTIVE REPORT. A PREVIOUS OCCURRENCE OF THIS ISSUE WAS REPORTED AS MFR REPORT #: 3011233554-2019-00006 AND ASSOCIATED WITH RECALL 30112335544-11-18-19-003-C. THIS ISSUE WAS IN THE RECALL BUT WAS NOT INDIVIDUALLY REPORTED AS AN MDR. THE RECALL FOR THIS ISSUE HAS BEEN COMPLETED AND A SOFTWARE PATCH WAS RELEASED. VIEWRAY HAS NOT RECEIVED A CUSTOMER REPORT OF INCORRECT DOSE APPLICATION OR SIGNIFICANT DELAY IN TREATMENT DUE TO THIS ISSUE.

Description of Event or Problem · 0

A USER OF VIEWRAY'S MRIDIAN LINAC SYSTEM REPORTED AN OCCURRENCE IN THE PLANNING/FUSION WORKFLOW WHERE: 1. MANUAL TRANSLATION SHIFT CANNOT BE APPLIED CORRECTLY TO SECONDARY DOSE VOLUME WHEN SECONDARY DOSE AND SECONDARY CT COMES FROM OTHER VENDOR A. MANUALLY ALIGN SECONDARY CT AND PRIMARY MRI. B. APPLY THE SAME TRANSLATION VALUE TO THE SECONDARY DOSE WHICH IS ASSOCIATED TO THE CT. C. OBSERVED THAT SECONDARY DOSE VOLUME IS NOT ALIGNED WITH PRIMARY MRI AND REGISTERED CT. 2. MANUAL TRANSLATION SHIFT CANNOT BE APPLIED CORRECTLY TO SECONDARY DOSE VOLUME WHEN SECONDARY DOSE AND SECONDARY MRI GENERATED FROM MRIDIAN SYSTEM BUT AT A DIFFERENT FOV COMPARED TO PRIMARY MRI. A. MANUALLY ALIGN SECONDARY MRIDIAN MRI AND PRIMARY MRIDIAN MRI. B. APPLY THE SAME TRANSLATION VALUE TO THE SECONDARY DOSE WHICH IS ASSOCIATED TO THE SECONDARY MRIDIAN MRI. C. OBSERVED THAT SECONDARY DOSE VOLUME IS NOT ALIGNED WITH PRIMARY MRI AND REGISTERED MRI.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1810282 MRIDIAN LINAC RADIATION THERAPY IYE VIEWRAY, INC. 20000 10868029000305

Patients

Seq Age Sex Outcome Treatment
1 Unknown