FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE

MDR report key: 12913017 · Received December 1, 2021

Report

Report Number
3003152976-2021-00791
Event Type
Malfunction
Date Received
December 1, 2021
Date of Event
November 1, 2021
Report Date
December 29, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: PHOTO RECEIVED FOR INVESTIGATION, UPON VISUAL INSPECTION OF THE PICTURES RECEIVED IT CAN BE CONFIRMED LEAKAGE PAST THE STOPPER IN THE SYRINGE SHOWN. NO DAMAGE IN ANY OF THE VISIBLE PARTS OF THE SYRINGE CAN BE SEEN. STOPPER IS CORRECTLY ASSEMBLED IN THE DEVICE. TEN RETAINED SAMPLES FROM THE SAME LOT WERE EVALUATED, NO DAMAGE CAN BE OBSERVED. STOPPER IS CORRECTLY IN ALL OF THE SYRINGES. FUNCTIONAL TEST WAS PERFORMED, NO LEAKAGE OBSERVED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2010110, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED ISSUE. BASED ON THE QUALITY TEAM'S INVESTIGATION, WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME.

Additional Manufacturer Narrative · 0

INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1808126 BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 2010110

Patients

Seq Age Sex Outcome Treatment
1 Unknown