FDA Adverse Event Malfunction Summary report: N

REBEL

MDR report key: 12910310 · Received December 1, 2021

Report

Report Number
2134265-2021-15070
Event Type
Malfunction
Date Received
December 1, 2021
Date of Event
September 6, 2021
Report Date
December 1, 2021
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
MAF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATED BY MFR: RETURNED PRODUCT CONSISTED OF A REBEL BMS BALLOON CATHETER. THE DEVICE WAS VISUALLY AND MICROSCOPICALLY EXAMINED. THERE WAS CONTRAST IN THE INFLATION LUMEN AND BLOOD IN THE GUIDEWIRE LUMEN. THE RAPID PORT EXCHANGE WAS DAMAGED. THE STENT WAS FOUND MISPLACED WITHIN 1MM DISTAL FROM THE PROXIMAL MARKERBAND. THE STENT BRACKET MOST PROXIMAL FROM THE TIP OF THE DEVICE WAS PULLED UP AWAY FROM THE BALLOON. THE BALLOON WAS TIGHTLY FOLDED WITH THE STENT ATTACHED, AND THE TIP OF THE DEVICE WAS DAMAGED. PRODUCT ANALYSIS CONFIRMED THE REPORTED EVENT, AS THERE WAS RAPID PORT EXCHANGE AND TIP DAMAGE LIKELY CONTRIBUTING TO THE REPORTED CROSSING DIFFICULTY, WHICH COULD NOT BE CONFIRMED BECAUSE THE CLINICAL CIRCUMSTANCES COULD NOT BE REPLICATED. THE STENT TO THE DEVICE WAS ALSO DAMAGED. THIS REFLECTS THE DEVICE GETTING DAMAGED UPON REMOVAL AND CAN ALSO CONTRIBUTE TO THE REPORTED CROSSING DIFFICULTY.

Description of Event or Problem · 0

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 23 NOVEMBER 2021. IT WAS REPORTED THAT FAILURE TO CROSS A LESION OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 24 X 3.00 REBEL STENT WAS ADVANCED, BUT COULD NOT PASS THROUGH THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS RESULTED IN RELATION TO THIS EVENT. HOWEVER, DEVICE RETURNED AND ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1808448 REBEL STENT, CORONARY MAF BOSTON SCIENTIFIC CORPORATION 10289 0025939740

Patients

Seq Age Sex Outcome Treatment
1 Unknown