BD ULTRA-FINE¿ NANO¿ PEN NEEDLE

MDR report key: 12906088 · Received November 30, 2021

Report

Report Number
9616656-2021-01463
Event Type
Malfunction
Date Received
November 30, 2021
Date of Event
November 1, 2021
Report Date
January 13, 2022
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
UDI-DI
00382903205509
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A COMPLAINT LOT HISTORY CHECK WAS PERFORMED ON LOT # 1040138 FOR NEEDLE CLOG. THIS IS THE 2ND. RELATED COMPLAINT FOR NEEDLE CLOG ON LOT # 1040138. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.9. DEVICE AVAILABLE FOR EVAL?: YES. D.9. RETURNED TO MANUFACTURER ON: 1/6/2022. H.6. INVESTIGATION: CUSTOMER RETURNED A SINGLE USED 4MM, 32 GAUGE PEN NEEDLE. NO TEARDROP LABEL WAS AVAILABLE FOR LOT CONFIRMATION. THE PEN NEEDLE WAS VISUALLY INSPECTED PRIOR TO TESTING. THE PEN NEEDLE WAS FOUND TO HAVE BEEN BENT ON THE NON-PATIENT SIDE OF THE HUB¿S NEEDLE CANNULA. THIS DAMAGE WOULD PREVENT TESTING FOR CLOGS SINCE THE NEEDLE CANNOT PROPERLY ATTACH TO THE PEN. AS A RESULT, THE PEN NEEDLES APPEAR TO BE CLOGGED DURING USE. BASED ON THE DAMAGE PRESENT, THE NEEDLE BENDING MAY HAVE OCCURRED ACCIDENTALLY WHILE THE USER WAS PREPARING THE PEN NEEDLE FOR USE, POTENTIALLY IF THE PEN WAS NOT PROPERLY ALIGNED WITH THE CANNULA. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. BASED ON THE SAMPLES RECEIVED, BD FOUND THAT THE PEN NEEDLE HAD BECOME BENT ON THE NON-PATIENT SIDE. THIS DAMAGE COULD RESEMBLE THE NEEDLE BEING CLOGGED AS A RESULT OF INSULIN NOT PASSING THROUGH THE NEEDLE. THE ROOT CAUSE OF THE CANNULA BENDING ON THE NON-PATIENT SIDE APPEARS TO BE ACCIDENTAL DAMAGE FROM STRESSES CAUSED BY THE USER DURING ROUTINE USE. THIS WOULD HAVE ALSO GIVEN THE IMPRESSION THAT THE NEEDLE WAS CLOGGED SINCE INSULIN WOULD BE UNABLE TO PASS THROUGH THE CANNULA.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD ULTRA-FINE¿ NANO¿ PEN NEEDLE WAS UNABLE TO PRIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER REPORTED NO INSULIN FLOW DURING PRIMING. DATE OF EVENT : UNKNOWN. SAMPLES : AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD ULTRA-FINE¿ NANO¿ PEN NEEDLE WAS UNABLE TO PRIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER REPORTED NO INSULIN FLOW DURING PRIMING. DATE OF EVENT : UNKNOWN. SAMPLES : AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD ULTRA-FINE¿ NANO¿ PEN NEEDLE WAS UNABLE TO PRIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER REPORTED NO INSULIN FLOW DURING PRIMING. DATE OF EVENT : UNKNOWN. SAMPLES : AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1800832 BD ULTRA-FINE¿ NANO¿ PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 320550 1040138 00382903205509

Patients

Seq Age Sex Outcome Treatment
1 Unknown