FDA Adverse Event Malfunction Summary report: N

REGULATED SOFTWARE APPLICATION (RSA), VERSION 2.0

MDR report key: 12901067 · Received November 30, 2021

Report

Report Number
3016690356-2021-00001
Event Type
Malfunction
Date Received
November 30, 2021
Date of Event
November 24, 2019
Report Date
August 12, 2021
Manufacturer
ARC-ONE SOLUTIONS, LLC
Product Code
MMH
UDI-DI
0860006793708
PMA / PMN Number
BK200512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

REGULATED SOFTWARE APPLICATION (RSA), V2.0 IS USED BY A SINGLE CORPORATE ENTITY. THIS SINGLE ENTITY WAS NOTIFIED OF THE ISSUE, WORKAROUND, AND SPECIFICALLY IMPACTED BLOOD PRODUCTS ON (B)(6) 2021. A QUERY OF ALL APPLICABLE BLOOD PRODUCT RECORDS IDENTIFIED 12 LABELED AS-3 RED BLOOD CELLS LEUKOCYTES REDUCED PRODUCTS THAT WERE OUT OF STORAGE FOR 34 MINUTES (4 MINUTES PAST THE CLIENT DEFINED SPECIFICATION) AND SHIPPED ON (B)(6) 2019. INFORMATION OF THESE UNITS WAS PROVIDED TO THE CLIENT. THE CLIENT HAS PROVIDED EMAIL NOTIFICATION THAT COMMUNICATION WAS RECEIVED AND ARE FURTHER EVALUATING THESE UNITS FOR FURTHER ACTION. TO DATE, NO REPORTS OF DEATH OR SERIOUS INJURY HAVE BEEN IDENTIFIED FOR THIS ISSUE. A SOFTWARE FIX IS IN PROGRESS. DETAILS OF COMPLETION AND RELEASE WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

A SOFTWARE MODIFICATION WAS PERFORMED, VERIFIED, AND VALIDATED. THE SCOPE OF TESTING INCLUDED THE SOFTWARE MODIFICATION ITSELF, ALL POTENTIALLY IMPACTED AREAS OF THE SOFTWARE, AND REGRESSION TESTING TO ENSURE THAT THE FIX WAS EFFECTIVE AND DID NOT ADVERSELY AFFECT THE SOFTWARE. THIS SOFTWARE MODIFICATION WAS RELEASED ON 12/10/2021 AS V2.0.2.1. WHILE THE SOFTWARE IS READY FOR IMPLEMENTATION, THE SOLE USER GROUP WAS INFORMED OF THE CORRECTIVE VERSION OF THE SOFTWARE BEING AVAILABLE. THEY REQUESTED DELAY UNTIL THEY ARE ADEQUATELY PREPARED TO INSTALL. AS SUCH, ARC-ONE SOLUTIONS WILL UPLOAD A FOLLOW-UP REPORT WHEN THE SOLE USER GROUP IS RUNNING ON THE CORRECTIVE VERSION (V2.0.2.1).

Additional Manufacturer Narrative · 0

CURRENTLY, THE BECS IS USED ONLY BY A SINGLE CORPORATE ENTITY. ON (B)(6) 2022, THE PATCH VERSION OF THE SOFTWARE WAS INSTALLED BY OUR SINGLE CLIENT. AS SUCH, ALL USERS AFFECTED BY THE ISSUE ARE CURRENTLY WORKING WITH THE UPDATED SOFTWARE APPLICATION.

Description of Event or Problem · 0

ARC-ONE SOLUTIONS RECEIVED INFORMATION REGARDING REGULATED SOFTWARE APPLICATION (RSA), V2.0 NOTING POTENTIAL MISCALCULATION OF STORAGE RECORDS FOR PRODUCTS THAT WENT THROUGH THE IRRADIATION PROCESS BE BEFORE MIDNIGHT AND FINISH AFTER MIDNIGHT. THIS RARE SCENARIO RESULTS IN A MISCALCULATION WHERE "TIME TAKEN OUT OF STORAGE" IS INTERPRETED BY THE SOFTWARE AS A NEGATIVE VALUE CAUSING A FAILURE TO TRIGGER DOWNSTREAM EFFECTS SUCH AS APPLICATION OF A QUARANTINE IDENTIFICATION AND PREVENTION OF UNIT DISTRIBUTION. DOWNSTREAM EFFECTS HAVE BEEN CONFIRMED TO WORK AGAIN AFTER AN ADDITIONAL 24-HOUR TIME FRAME IS PASSED, RESULTING IN A POSITIVE "TIME TAKEN OUT OF STORAGE" VALUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1800045 REGULATED SOFTWARE APPLICATION (RSA), VERSION 2.0 BLOOD ESTABLISHMENT COMPUTER SOFTWARE AND ACCESSORIES MMH ARC-ONE SOLUTIONS, LLC 2.0.1.0 0860006793708

Patients

Seq Age Sex Outcome Treatment
1 Unknown