CUSTOM ULTRASONICS
Report
- Report Number
- 2523209-2009-00001
- Event Type
- Other
- Date Received
- January 13, 2009
- Date of Event
- December 15, 2008
- Report Date
- January 12, 2009
- Manufacturer
- CUSTOM ULTRASONICS, INC.
- Product Code
- FLG
- PMA / PMN Number
- K983017
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
OUR SERVICE TECHNICIAN WAS DISPATCHED TO THE FACILITY UPON LEARNING OF THE INCIDENT. TESTING AND INSPECTION OF THE 3 SYSTEM 83 REPROCESSORS REVEALED RANDOM AND ROUTINE STORAGE OF IRRIGATION BOTTLES AND DISINFECTANT BOWLS INSIDE THE DISINFECTANT RESERVOIR OF ALL 3 FACILITY OWNED CUSTOM ULTRASONICS DEVICES. WE HAVE CONCLUDED THE RESIDUAL DISINFECTANT WAS ABLE TO ENTER IRRIGATION BOTTLE AND LINES. UNLESS PROPERLY PROCESSED, RESIDUAL DISINFECTANT WILL REMAIN IN THE IRRIGATION BOTTLE LINES. FACILITY PERSONNEL COULD NOT PRODUCE ANY SOP FOR PROPER PROCESSING OF THE BOTTLES AND LINES. THESE ARE ACCESSORIES SUPPLIED BY ANOTHER MANUFACTURER AND NOT SUPPLIED BY CU. IT IS THEREFORE, ALSO LIKELY THAT DISINFECTANT SOLUTION WAS INTRODUCED INTO AT LEAST 4 PATIENTS AT THE TIME OF COLONOSCOPY VIA THE IRRIGATION BOTTLE AND OR ITS TUBING. DISINFECTION SOLUTION CAN CAUSE DAMAGE WHEN EXPOSED TO TISSUE. IN THE EVENT OF CONTACT WITH THE COLON, COLITIS CAN OCCUR. SERVICE TECHNICIAN COMPLETED THE EVALUATION AND TESTING OF ALL SYSTEM 83+9'S AT THIS FACILITY. STAFF WAS RETRAINED FOR PROPER USE OF THE REPROCESSOR AND WERE DIRECTED TO THE USER MANUAL REQUIREMENTS WHICH CAUTION AGAINST STORING ANYTHING IN THE DISINFECTANT RESERVOIR OF THE SYSTEM 83+9 OR INJURY TO PATIENTS COULD OCCUR. INFECTION CONTROL MANAGEMENT WERE URGED TO PROVIDE ON-SITE SURVEILLANCE OF CONTAINER STORAGE PRACTICES SO THAT NO RESIDUAL DISINFECTANT INJURY WOULD OCCUR. DISINFECTANT NOT SUPPLIED OR MARKETED BY CUSTOM ULTRASONICS, INC. THE 3 SYSTEM 83+9'S WERE FOUND TO BE IN PERFECT WORKING ORDER. WE DID RECOMMEND A ROUTINE PREVENTATIVE MAINTENANCE AND REPLACEMENT OF SCOPE ADAPTERS WHICH WERE FOUND TO BE OLD.
POSSIBLE COLITIS CONDITION REPORTED FOR 4 PATIENTS RECEIVING COLONOSCOPIES IN 2008, FOLLOWING COLONOSCOPE REPROCESSING USING CUSTOM ULTRASONICS SYSTEM 83+9.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CUSTOM ULTRASONICS | SYSTEM 83+9 WASHER /DISINFECTOR | FLG | CUSTOM ULTRASONICS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | & FLUSH BOWLS ROUTINELY USED.| COLONOSCOPE ACCESSORY IRRIGATION SUPPLY BOTTLES |