FDA Adverse Event Injury Summary report: N

BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 36 SIZE M 0

MDR report key: 12890288 · Received November 30, 2021

Report

Report Number
3005180920-2021-00961
Event Type
Injury
Date Received
November 30, 2021
Date of Event
November 12, 2021
Report Date
November 30, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805219
PMA / PMN Number
K080885
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 15.NOVEMBER.2021: LOT 2104424: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-07-2021. EXPIRATION DATE: 2026-06-22 NO ANOMALIES WERE FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENTS. ADDITIONAL IMPLANT INVOLVED: LINER: MPACT 01.32.3652HCT FLAT PE HC LINER Ø36/G (K103721) LOT. 2100163. BATCH REVIEW PERFORMED ON 15.NOVEMBER.2021: LOT 2100163: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 1-FEB-2021. EXPIRATION DATE: 2026-01-19. NO ANOMALIES WERE FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENTS.

Description of Event or Problem · 0

THE PATIENT CAME IN 3 DAYS AFTER THE PRIMARY SURGERY REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER. THE SURGEON REVISED THE CUP, HEAD, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE SURGEON IMPLANTED A LARGER DM CUP WITH MORE ANTEVERSION PUT IN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1794634 BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 36 SIZE M 0 METAL BALL HEAD JDI MEDACTA INTERNATIONAL SA 01.25.031 2104424 07630030805219

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention