FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 12883673 · Received November 29, 2021

Report

Report Number
2647876-2021-00300
Event Type
Malfunction
Date Received
November 29, 2021
Date of Event
November 2, 2021
Report Date
December 23, 2021
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY CUSTOMER REPORTED BLOOD UNDER THE SENSOR AND FALSE POSITIVE RESULT. TWO PHOTOS WERE RECEIVED WITH SENSOR ADHESION DEFECT. BD WAS UNABLE TO DUPLICATE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED WITH SATISFACTORY RESULTS. BATCH/SENSOR HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. BLOOD BACKGROUND WAS PERFORMED WITH SATISFACTORY RESULTS. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. SENSOR ADHESION SCRAPE TEST IS PERFORMED TO EACH SENSOR BATCH AS PART OF RELEASE CRITERIA. COMPLAINT IS CONFIRMED BASED ON PHOTOS RECEIVED. THE VISION SYSTEM IS CHALLENGED PRIOR EACH LOT. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. CORRECTIVE ACTIONS PREVENTIVE ACTIONS (CAPA) 2882676 WAS INITIATED TO FURTHER INVESTIGATE THESE TYPES OF COMPLAINTS AND DETERMINE ANY APPROPRIATE ACTIONS TO REDUCE THEIR OCCURRENCE. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), 2 FALSE POSITIVE RESULTS WERE OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING A GRAM STAIN AND THE RESULT WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER HAD 2 BOTTLES THAT FLAGGED AS POSITIVE, BUT NOTHING WAS SEEN IN THE GRAM STAINS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), 2 FALSE POSITIVE RESULTS WERE OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING A GRAM STAIN AND THE RESULT WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER HAD 2 BOTTLES THAT FLAGGED AS POSITIVE, BUT NOTHING WAS SEEN IN THE GRAM STAINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1780166 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442023 1188574 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 Unknown