ALINITY C LACTATE DEHYDROGENASE REAGENT KIT
Report
- Report Number
- 3002809144-2021-00676
- Event Type
- Malfunction
- Date Received
- November 29, 2021
- Date of Event
- November 15, 2021
- Report Date
- January 12, 2022
- Manufacturer
- ABBOTT GMBH
- Product Code
- CFJ
- UDI-DI
- 00380740130244
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY C LACTATE DEHYDROGENASE RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, AND DEVICE HISTORY RECORD REVIEW. TRENDING REVIEW DID NOT IDENTIFY ANY TRENDS FOR THE ISSUE FOR THE PRODUCT. DEVICE HISTORY RECORD REVIEW FOR LOT 24339UN21 DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE UNDER REVIEW. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ALINITY C LACTATE DEHYDROGENASE REAGENT LOT 24339UN21 WAS IDENTIFIED.
PATIENT IDENTIFICATION: SIDS (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C LACTATE DEHYDROGENASE RESULTS FOR TWO SAMPLES. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE PROVIDED: 125 TO 220), UNITS OF MEASURE = U/L): (B)(6) 2021 SID (B)(6) INITIAL RESULT = 831, REPEATS = 286 AND 222. (B)(6) 2021 SID (B)(6) INITIAL RESULT = 201, REPEATS = 171, 157, AND 156. (B)(6) 2021 SID (B)(6) INITIAL RESULT = 268, REPEATS = 326 AND 313. THE RESULTS GENERATED WERE FROM AN AUTOMATIC 1:3 DILUTION THAT WAS PROGRAMED ON THE INSTRUMENT. THE SAMPLES WERE NOT RUN UNDILUTED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C LACTATE DEHYDROGENASE RESULTS FOR TWO SAMPLES. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE PROVIDED: 125 TO 220), UNITS OF MEASURE = U/L): (B)(6) 2021 SID (B)(6) INITIAL RESULT = 831, REPEATS = 286 AND 222 (B)(6) 2021 SID (B)(6) INITIAL RESULT = 201, REPEATS = 171, 157, AND 156 (B)(6) 2021 SID (B)(6) INITIAL RESULT = 268, REPEATS = 326 AND 313 THE RESULTS GENERATED WERE FROM AN AUTOMATIC 1:3 DILUTION THAT WAS PROGRAMED ON THE INSTRUMENT. THE SAMPLES WERE NOT RUN UNDILUTED. ADDITIONAL LABORATORY DATA WAS PROVIDED BY THE CUSTOMER: SID (B)(6): PLATELET COUNT = 800 SID (B)(6): PLATELET COUNT = 212 SID (B)(6): PLATELET COUNT = 559.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1781541 | ALINITY C LACTATE DEHYDROGENASE REAGENT KIT | NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE | CFJ | ABBOTT GMBH | 07P7420 | 24339UN21 | 00380740130244 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ALNTY C PROCESSING MODU, 03R67-01, (B)(6)| ALNTY C PROCESSING MODU, 03R67-01, (B)(6) |