FDA Adverse Event Malfunction Summary report: N

SUPPORT ARM 177

MDR report key: 12873588 · Received November 25, 2021

Report

Report Number
8010042-2021-02788
Event Type
Malfunction
Date Received
November 25, 2021
Report Date
November 25, 2021
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
IOY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

NO REPAIR WAS REQUESTED AND NO PARTS WERE RETURNED. THE SUPPORT ARM IS A CASTING AND IT MOST PROBABLY DEVELOPED A CRACK AT AN EARLIER OCCASION EITHER BY OVERLOADING OR IMPACT AND THIS LED TO THE REPORTED BREAKING. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUPPORT ARM BROKE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1773139 SUPPORT ARM 177 SUPPORT, ARM IOY MAQUET CRITICAL CARE AB SUPPORT ARM 177

Patients

Seq Age Sex Outcome Treatment
1 Unknown