BINAXNOW COVID-19 ANTIGEN SELF-TEST
Report
- Report Number
- 1221359-2021-03596
- Event Type
- Malfunction
- Date Received
- November 24, 2021
- Report Date
- December 6, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCABOROUGH
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE CORRECTED DATA AND ADDITIONAL INFORMATION.
INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 152599 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160 / LOT 152599 AND DEVICE PART NUMBER 195-430H / LOT 146941A. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 152599 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF TEST. THE CUSTOMER REPORTS THE INITIAL TEST GENERATED POSITIVE RESULTS. REPEAT TESTING WAS UNSUCCESSFUL. CONFIRMATION TESTING GENERATED NEGATIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1767122 | BINAXNOW COVID-19 ANTIGEN SELF-TEST | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM | QKP | ABBOTT DIAGNOSTICS SCABOROUGH | 152599 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female |