EASYPUMP® II
Report
- Report Number
- 9610825-2021-00483
- Event Type
- Malfunction
- Date Received
- November 24, 2021
- Date of Event
- October 15, 2021
- Report Date
- June 7, 2022
- Manufacturer
- B. BRAUN MELSUNGEN AG
- Product Code
- MEB
- UDI-DI
- 04046964448515
- PMA / PMN Number
- K081905
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER (B)(4). FAILURE ANALYSIS AND INVESTIGATION RESULTS DID NOT CONFIRM THE REPORTED ISSUE. UPON RECEIPT THE ACTUAL PUMP INVOLVED WAS VISUALLY AND FUNCTIONALLY INSPECTED. THREE (3) VOLUMETRICS OF 250ML/HR AND 25ML WERE PERFORMED AND TESTED IN SPEC. THE PUMP OPERATED AS INTENDED AND THE REPORTED FAILURE COULD NOT BE REPRODUCED. BASED ON THE RESULTS OF THE INVESTIGATION, NO CONCLUSIONS CAN BE MADE REGARDING THE CAUSE OF THE REPORTED EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4): EVENT 1: AS NO SAMPLE AND NO MEANINGFUL PICTURE WAS PROVIDED, PROPER INVESTIGATION ON MALFUNCTION COULD NOT BE PERFORMED. HENCE, THE COMPLAINT IS CONSIDERED AS NOT CONFIRMED. DEVICE HISTORY RECORD (DHR: REVIEWED THE DHR FOR BATCH 21B03GE231, THERE IS NO ABNORMALITY AND NO SUCH DEFECT DETECTED AT IN PROCESS AND AT FINAL CONTROL INSPECTION. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY / TRANSLATION OF USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN FRANCE: EVENT 1: "PUMP-FLOW RATE-FAST." "2 DIFFUSERS FILLED TO 48ML TO PASS OVER 24 HOURS PASSED IN HALF THE TIME, I.E. 12 HOURS. THE EVENT OCCURRED IN THE SAME PATIENT TWICE. THERE WERE NO CONSEQUENCES FOR THE PATIENT. NO SAMPLE AVAILABLE OCCURENCE DATE : (B)(6) 2021."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1766659 | EASYPUMP® II | ELASTOMERIC INFUSION PUMP | MEB | B. BRAUN MELSUNGEN AG | 4540010-02 | 21B03GE231 | 04046964448515 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |