BD SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL
Report
- Report Number
- 9614033-2021-00140
- Event Type
- Malfunction
- Date Received
- November 23, 2021
- Date of Event
- October 26, 2021
- Report Date
- December 13, 2021
- Manufacturer
- BECTON DICKINSON DE MEXICO
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: PHOTO RECEIVED FOR INVESTIGATION, UPON VISUAL INSPECTION AN ASSEMBLED NEEDLE WITH A LACK OF COMPONENT (SHIELD) WHICH COMES OUT OF THE BLISTER IN ITS UPPER PART, THAT IS, THROUGH THE PART OF THE FLANGE ON ONE SIDE OF THE PLUNGER IS OBSERVED. POSSIBLE ROOT CAUSE IS ASSOCIATED WITH LACK OF GUARD IN THE NEEDLE FEEDER. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE NEEDLE CAP WAS MISSING WITH A BD SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: DURING THE INSPECTION UPON RECEIPT, ONE NEEDLE WITHOUT A CAP (EXPOSED).
IT WAS REPORTED THAT THE NEEDLE CAP WAS MISSING WITH A BD SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: DURING THE INSPECTION UPON RECEIPT, ONE NEEDLE WITHOUT A CAP (EXPOSED).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1760955 | BD SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL | PISTON SYRINGE | FMF | BECTON DICKINSON DE MEXICO | 1197372 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |