BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 12862153 · Received November 23, 2021

Report

Report Number
3014704491-2021-00296
Event Type
Malfunction
Date Received
November 23, 2021
Date of Event
October 13, 2021
Report Date
November 8, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9023814. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM THE END CAP WAS LOOSE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "AFTER THE PATIENT WAS TREATED WITH A CLOSED INDWELLING NEEDLE INFUSION, THE WHITE CAP OF THE INDWELLING NEEDLE WAS FOUND TO HAVE FALLEN OFF WHEN THE CATHETER WAS SEALED AND FIXED FOR THE PATIENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1760587 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 9023814

Patients

Seq Age Sex Outcome Treatment
1 Unknown