FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 20ML LL 120/PKG

MDR report key: 12861483 · Received November 23, 2021

Report

Report Number
3003152976-2021-00774
Event Type
Malfunction
Date Received
November 23, 2021
Date of Event
October 26, 2021
Report Date
December 17, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION?: YES. D9: RETURNED TO MANUFACTURER ON: 11/29/2021. H6: INVESTIGATION: TWO SAMPLES AND PHOTOS RECEIVED FOR INVESTIGATION, UPON VISUAL INSPECTION OF THE SAMPLES RECEIVED IT CAN BE SEEN NO SCALE HAS BEEN PRINTED IN THE BARREL OF THE SYRINGES. A DEVICE HISTORY REVIEW WAS PERFORMED AND FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF LOT 2102013, ALL PRODUCT WAS MANUFACTURED ACCORDING TO SPECIFICATION. POSSIBLE ROOT CAUSE, DUE TO A FAILURE IN THE TRIGGER THAT GUIDES THE BARREL TO BE CORRECTLY POSITIONED IN THE MOMENT OF SCALE TRANSFERENCE IN THE MARKING MACHINE. THIS FAILURE CAUSED A MISALIGNMENT CAUSING SCALE RESULTED COMPRESSED AND NOT PRINTED.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE 20ML LL 120/PKG HAD SCALE MARKING MISSING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE WOULD LIKE TO INFORM YOU THAT WE HAVE FOUND 20 ML LUER LOCK SYRINGES (REF (B)(4)) WITHOUT GRADUATION".

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE 20ML LL 120/PKG HAD SCALE MARKING MISSING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE WOULD LIKE TO INFORM YOU THAT WE HAVE FOUND 20 ML LUER LOCK SYRINGES (REF (B)(4)) WITHOUT GRADUATION ".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1757037 BD SYRINGE 20ML LL 120/PKG PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2102013

Patients

Seq Age Sex Outcome Treatment
1 Unknown