BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM
Report
- Report Number
- 3014704491-2021-00294
- Event Type
- Malfunction
- Date Received
- November 22, 2021
- Date of Event
- October 24, 2021
- Report Date
- November 4, 2021
- Manufacturer
- BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0231256. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WILL CONTINUE TO MONITOR THIS ISSUE.
IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " FLUID EXUDATION WAS FOUND AT THE PRN"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1753882 | BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU | 0231256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |