FDA Adverse Event Injury Summary report: N

MICROWAVE ENDOMETRIAL ABLATION (MEA) SYSTEM

MDR report key: 1285762 · Received January 9, 2009

Report

Report Number
9710493-2008-00008
Event Type
Injury
Date Received
January 9, 2009
Date of Event
August 5, 2008
Report Date
January 8, 2009
Manufacturer
MICROSULIS MEDICAL LTD.
Product Code
MNB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL DEVICE MANUFACTURE DATE: SYSTEM 06/2004. THE MEA SYSTEM RECORDS DATA FOR EACH PATIENT TREATMENT PROCEDURE, INCLUDING SYSTEM PARAMETERS AND PATIENT DATA ENTERED BY THE PHYSICIAN. THE COMPANY ANALYZED THE MEA SYSTEM'S TREATMENT DATA FILE ASSOCIATED WITH THIS EVENT. THE CONCLUSION OF THE ANALYSIS WAS THAT THE MEA SYSTEM WAS FUNCTIONING WITHIN OPERATING SPECIFICATIONS AT THE TIME OF THE TREATMENT, AND NO MALFUNCTIONS OR PERFORMANCE DEVIATIONS WERE PRESENT. THE MEA IFU CONTRAINDICATES TREATMENT IN A PATIENT WITH A MYOMETRIAL WALL THICKNESS LESS THAN 10 MM. THE IFU PROVIDES DIRECTION ON MEASURING THE MYOMETRIAL WALL THICKNESS WITH THE USE OF TRANSVAGINAL ULTRASOUND.

Description of Event or Problem · 1

PATIENT WAS TREATED WITH A REUSABLE APPLICATOR. IMMEDIATELY FOLLOWING THE MEA PROCEDURE, A SCHEDULED LAPAROSCOPIC TUBAL LIGATION WAS PERFORMED. AT THE TIME OF THE LAPAROSCOPY, A THERMAL INJURY 2X1 CM WAS NOTED ON THE ANTERIOR SURFACE OF THE UTERUS BUT NO APPARENT INJURY TO THE BOWEL WAS IDENTIFIED AT THIS TIME. THE PATIENT WAS DISCHARGED WITH ANTIBIOTICS THE FOLLOWING DAY, HOWEVER, WAS READMITTED 4 DAYS POST MEA WITH COMPLAINT OF ABDOMINAL PAIN. SEVEN DAYS POST MEA, LAPAROSCOPY WAS PERFORMED AND SHOWED "A BREAKDOWN OF THE THERMAL INJURY WITH A FISTULA FROM THE UTERUS TO THE PERITONEUM." THE PATIENT SUBSEQUENTLY UNDERWENT LAPAROTOMY, HYSTERECTOMY AND OVERSEWING OF A 5MM BOWEL PERFORATION. THE PATIENT HAD AN UNEVENTFUL RECOVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROWAVE ENDOMETRIAL ABLATION (MEA) SYSTEM MICROWAVE ENDOMETRIAL ABLATION DEVICE MNB MICROSULIS MEDICAL LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention