FDA Adverse Event Injury Summary report: N

TRIUMPH-1

MDR report key: 128497 · Received October 28, 1997

Report

Report Number
128497
Event Type
Injury
Date Received
October 28, 1997
Date of Event
October 11, 1997
Report Date
October 28, 1997
Manufacturer
HORIZON MEDICAL PROD
Product Code
LJT
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

A TRIUMPH-1 VASCULAR ACCESS PORT WAS IMPLANTED IN THE RT CHEST ON 10/10/97. MEDICATIONS INFUSED WITHOUT DIFFICULTY UNTIL SWELLING AT THE SITE AND A WET DRESSING WERE REPORTED ON 10/11/1997. A CHEST AP-PA-X-RAY SHOWED CATHETER DISLODGEMENT TO THE RIGHT ATRIUM. A FLUORO GUIDED RETRIEVAL OF THE CATHETER FRAGMENT WAS SUCCESSFUL ON 10/11/1997. THE PATIENT WAS TAKEN TO SURGERY ON 10/14/1997 FOR INSERTION OF A NEW VASCULAR ACCESS PORT AND REMOVAL OF THE OLD PORT. ON CHECKING THE OLD PORT; THE SNAPPING MECHANISM REPORTEDLY WOULD NOT LOCK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIUMPH-1 Implant PORT LJT HORIZON MEDICAL PROD * 500133

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention