FDA Adverse Event
Injury
Summary report: N
TRIUMPH-1
MDR report key: 128497
·
Received October 28, 1997
Report
- Report Number
- 128497
- Event Type
- Injury
- Date Received
- October 28, 1997
- Date of Event
- October 11, 1997
- Report Date
- October 28, 1997
- Manufacturer
- HORIZON MEDICAL PROD
- Product Code
- LJT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
A TRIUMPH-1 VASCULAR ACCESS PORT WAS IMPLANTED IN THE RT CHEST ON 10/10/97. MEDICATIONS INFUSED WITHOUT DIFFICULTY UNTIL SWELLING AT THE SITE AND A WET DRESSING WERE REPORTED ON 10/11/1997. A CHEST AP-PA-X-RAY SHOWED CATHETER DISLODGEMENT TO THE RIGHT ATRIUM. A FLUORO GUIDED RETRIEVAL OF THE CATHETER FRAGMENT WAS SUCCESSFUL ON 10/11/1997. THE PATIENT WAS TAKEN TO SURGERY ON 10/14/1997 FOR INSERTION OF A NEW VASCULAR ACCESS PORT AND REMOVAL OF THE OLD PORT. ON CHECKING THE OLD PORT; THE SNAPPING MECHANISM REPORTEDLY WOULD NOT LOCK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIUMPH-1 Implant | PORT | LJT | HORIZON MEDICAL PROD | * | 500133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention |