FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM

MDR report key: 12849106 · Received November 19, 2021

Report

Report Number
3014704491-2021-00285
Event Type
Malfunction
Date Received
November 19, 2021
Date of Event
October 1, 2021
Report Date
November 3, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1020983. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WILL CONTINUE TO MONITOR THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM LEAKED AT THE CONNECTION SITE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " AFTER FIXATION, IT WAS FOUND THAT THE LEAKING FLUID AT THE NEEDLE HANDLE OF THE CLOSED VENOUS INDWELLING NEEDLE COULD NOT BE CORRECTED"

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1739306 BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1020983

Patients

Seq Age Sex Outcome Treatment
1 Unknown