FDA Adverse Event Injury Summary report: N

ENDO GIA

MDR report key: 12846186 · Received November 19, 2021

Report

Report Number
1219930-2021-04827
Event Type
Injury
Date Received
November 19, 2021
Date of Event
November 3, 2021
Report Date
March 25, 2022
Manufacturer
COVIDIEN LP LLC NORTH HAVEN
Product Code
GDW
UDI-DI
10884523003215
PMA / PMN Number
K111825
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE DEVICE WAS APPLIED ON OVER INDICATED TISSUE. IT WAS REPORTED THAT THE STAPLES WERE MISSING FROM THE STAPLE LINE AFTER FIRING AND UNEXPECTED BLEEDING OCCURRED ALONG THE STAPLE LINE. THE STAPLES DID NOT FORM AS EXPECTED. THE REPORTED ISSUES WERE CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THIS ISSUE CAN OCCUR UNDER THE FOLLOWING CONDITIONS. APPLICATION OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE. APPLICATION WITH AN OBSTACLE INCORPORATED IN THE JAWS. IN ANY OF THESE CIRCUMSTANCES, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY AND TISSUE MAY NOT BE FULLY TRANSECTED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FAILURE TO COMPLETELY FIRE THE RELOAD WILL RESULT IN AN INCOMPLETE CUT AND/OR INCOMPLETE STAPLE FORMATION, WHICH MAY RESULT IN POOR HEMOSTASIS AND/OR LEAKAGE. ALWAYS INSPECT THE TISSUE THICKNESS AND SELECT AN APPROPRIATE STAPLE SIZE PRIOR TO APPLICATION OF THE ENDO GIA¿ ULTRA UNIVERSAL SHORT, ENDO GIA¿ ULTRA UNIVERSAL OR ENDO GIA¿ ULTRA UNIVERSAL XL STAPLER. OVERLY THICK OR THIN TISSUE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION. WHEN POSITIONING THE STAPLER ON THE APPLICATION SITE, ENSURE THAT NO OBSTRUCTIONS, SUCH AS CLIPS, ARE INCORPORATED INTO THE INSTRUMENT JAWS. FIRING OVER AN OBSTRUCTION MAY RESULT IN INCOMPLETE CUTTING ACTION AND/OR IMPROPERLY FORMED STAPLES. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC LOWER ANTERIOR RESECTION, THE STAPLES DID NOT FORM A PROPER 'B-SHAPED', AND THE STAPLE LINE WAS INCOMPLETE PROXIMALLY. THE JAWS OF THE DEVICE LOCKED ON TISSUE, BUT IT WAS REMOVED BY PULLING THE TRIGGER. THERE WAS OOZING AT THE STAPLE LINE. ANOTHER STAPLER AND RELOAD WERE USED TO COMPLETE THE CASE. THE SURGICAL TIME WAS EXTENDED FOR 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1741148 ENDO GIA STAPLE, IMPLANTABLE GDW COVIDIEN LP LLC NORTH HAVEN EGIA60AMT N9K0744KY 10884523003215

Patients

Seq Age Sex Outcome Treatment
1 53 YR Female Required Intervention