FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12841817 · Received November 19, 2021

Report

Report Number
9610877-2021-50423
Event Type
Malfunction
Date Received
November 19, 2021
Date of Event
November 9, 2021
Report Date
November 19, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO AN EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE THE ANGLE WIRE HARD TO MOVE; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1745027 PENTAX VIDEO COLONOCOPE - I10 STANDARD FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC38-I10F2

Patients

Seq Age Sex Outcome Treatment
1 Unknown