FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12841817
·
Received November 19, 2021
Report
- Report Number
- 9610877-2021-50423
- Event Type
- Malfunction
- Date Received
- November 19, 2021
- Date of Event
- November 9, 2021
- Report Date
- November 19, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO AN EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE THE ANGLE WIRE HARD TO MOVE; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1745027 | PENTAX | VIDEO COLONOCOPE - I10 STANDARD | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC38-I10F2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |