FDA Adverse Event Injury Summary report: N

3M¿ STERI-VAC¿ STERILIZER, 1 DOOR GS8X-1D

MDR report key: 12840212 · Received November 18, 2021

Report

Report Number
2110898-2021-00068
Event Type
Injury
Date Received
November 18, 2021
Date of Event
October 21, 2021
Report Date
January 25, 2022
Manufacturer
3M HEALTH CARE
Product Code
FLF
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE STERILIZER WAS EXAMINED AND NO ISSUES WERE FOUND. THE 3M STERI-GAS EO CARTRIDGE LIKELY HAD A PRECONDITIONING FAILURE IN THE STERILIZER. THIS CARTRIDGE FAILURE RESULTED IN AN ERROR ON THE STERILIZER, AND ETHYLENE OXIDE WAS IN THE STERILIZATION CHAMBER WITHOUT THE CARTRIDGE HAVING BEEN PUNCTURED. 3M¿ STERI-VAC¿ STERILIZER, 1 DOOR GS8X-1D2 IS AN INDUSTRIAL DESIGN THAT IS NOT REGULATED AS A MEDICAL DEVICE. THIS REPORT WAS RECEIVED FROM AN INDUSTRIAL CUSTOMER AND NOT FROM A HEALTHCARE FACILITY.

Additional Manufacturer Narrative · 0

PATIENT IDENTIFIER, WEIGHT, ETHNICITY, RACE: NOT PROVIDED. THE STERILIZER WAS EXAMINED AND NO ISSUES WERE FOUND. THE 3M STERI-GAS EO CARTRIDGE LIKELY HAD A PRECONDITIONING FAILURE IN THE STERILIZER. THIS CARTRIDGE FAILURE RESULTED IN AN ERROR ON THE STERILIZER, AND ETHYLENE OXIDE WAS IN THE STERILIZATION CHAMBER WITHOUT THE CARTRIDGE HAVING BEEN PUNCTURED.

Description of Event or Problem · 0

A LABORATORY EMPLOYEE EXPERIENCED A LIGHTHEADEDNESS AND HEADACHE AFTER ALLEGED EXPOSURE TO ETHYLENE OXIDE FROM A 3M STERI-GAS EO CARTRIDGE THAT OCCURRED AFTER A 3M STERI-VAC STERILIZER, 1 DOOR GS8X-1D DOOR WAS OPENED. AN ERROR MESSAGE APPEARED ON THE STERILIZER (E108) THE NEXT MORNING AFTER THE STERILIZER RAN OVERNIGHT AND AN EMPLOYEE OPENED THE STERILIZER DOOR. AN ETHYLENE OXIDE ALARM SOUNDED. THE EMPLOYEE WAS SENT TO URGENT CARE, WHERE VITAL SIGNS WERE CHECKED AND FOUND TO BE NORMAL. NO MEDICATION OR OTHER TREATMENT WAS PROVIDED. A FOLLOW-UP APPOINTMENT WAS SCHEDULED FOR THE FOLLOWING DAY AND THE SYMPTOMS HAD SUBSIDED. NO ISSUES WERE FOUND WITH THE STERILIZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1737547 3M¿ STERI-VAC¿ STERILIZER, 1 DOOR GS8X-1D STERILIZER, ETHYLENE-OXIDE GAS FLF 3M HEALTH CARE GS8X-1D

Patients

Seq Age Sex Outcome Treatment
1 22 YR Male Other