BD BBL¿ COLUMBIA AGAR PLATE WITH 5% SHEEP BLOOD
Report
- Report Number
- 9680577-2021-00082
- Event Type
- Malfunction
- Date Received
- November 18, 2021
- Date of Event
- October 25, 2021
- Report Date
- January 13, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- KZI
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: THIS STATEMENT IS TO SUMMARIZE FINDINGS ON THE RECENT COMPLAINT AGAINST BDTM COLUMBIA AGAR WITH 5 % SHEEP BLOOD, CATALOG NUMBER 254071, LOT NUMBER 1243759 WITH RESPECT TO CONTAMINATION. EVENT DESCRIPTION: SEVERAL PLATED MEDIA OF LOT 1243759 WERE FOUND CONTAMINATED. COMPLAINT HISTORY REVIEW: THE COMPLAINT TREND WAS REVIEWED FOR A PERIOD COVERING 12 MONTHS. THE AMOUNT OF REGISTERED COMPLAINTS REGARDING CONTAMINATION OF LOT 1243759 REACHED AN INTERNAL ACTION LIMIT. A TREND COULD BE IDENTIFIED. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES FROM VALIDATED PRODUCTION PROCESSES. HOWEVER, QC RELEASE TESTING OF ONE SUBLOT FOR LOT 1243759 SHOWED SUBLIMINAL CONTAMINATION. THE INTERNAL ACCEPTANCE QUALITY LIMIT (AQL) FOR STERILITY WAS NOT VIOLATED. SAMPLE ANALYSIS: ACCORDING TO CUSTOMER SERVICE RETURN SAMPLES WERE PROVIDED BY THE CUSTOMER. UNFORTUNATELY THEY WERE LOST IN TRANSFER. PICTURE SAMPLES, PROVIDED BY THE CUSTOMER, SHOW PLATES WITH VISIBLE MICROBIAL GROWTH PRESENT. RETAIN SAMPLES OF LOT 1243759 WERE ANALYZED AND SHOWED NO SIGN OF CONTAMINATION. EVALUATION RESULTS: BASED ON THE INTERNAL INVESTIGATION AND THE PICTURE SAMPLES PROVIDED, THIS COMPLAINT CAN BE CONFIRMED FOR CONTAMINATION. IN ADDITION, A COMPLAINT TREND FOR CATALOG NUMBER 254071 WAS IDENTIFIED. THEREFORE, AN INTERNAL TEAM WAS FORMED FOR INVESTIGATION PURPOSES. A DEFINITE ROOT CAUSE COULD NOT BE IDENTIFIED. WE ARE SORRY FOR THE INCONVENIENCE. INVESTIGATION CONCLUSION: ASEPTIC MANUFACTURING PROCESSES ARE UNABLE TO GUARANTEE STERILITY OF THE PRODUCT. SINCE A 100 % INSPECTION IS NOT POSSIBLE FOR ASEPTIC PRODUCTS, STERILITY TESTING IS CARRIED OUT ON THE BASIS OF A REPRESENTATIVE SAMPLE. THEREFORE, OCCASIONAL CONTAMINATION CANNOT BE PREVENTED. HOWEVER, A TREND WAS IDENTIFIED. A TEAM WAS FORMED TO IMPROVE INTERNAL PROCESSES AND REDUCE THE OCCURRENCE RATE OF SUBLIMINAL CONTAMINATION. EFFECTIVENESS CHECKS INDICATE A SIGNIFICANT IMPROVEMENT TO THE SITUATION CAUSED BY THE TRIGGERED ACTIONS. BD WILL CONTINUE TO MONITOR INCOMING COMPLAINTS FOR SIMILAR DEFECT TYPES. WE WOULD SUGGEST TO SET ASIDE, AND NOT USE, ANY PREPARED PLATED MEDIA THAT DOES NOT MEET THE APPEARANCE AND PERFORMANCE SPECIFICATION AS IT IS DESCRIBED ON THE BD CERTIFICATE OF ANALYSIS. THIS IS CONSISTENT WITH INDUSTRY RECOMMENDATIONS FOR INSPECTION OF CULTURE MEDIA PRIOR TO USE (E.G. ¿GOOD PRACTICES FOR PHARMACEUTICAL MICROBIOLOGY LABORATORIES¿, WHO TECHNICAL REPORT SERIES, NO. 961, 2011, ANNEX 2; CHAPTER <1117> ¿MICROBIOLOGY BEST LABORATORY PRACTICES¿ THE UNITED STATES PHARMACOPEIA; AND THE DIFCO & BBL MANUAL). H3 OTHER TEXT : SEE H.10.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD BBL¿ COLUMBIA AGAR PLATE WITH 5% SHEEP BLOOD WAS FOUND CONTAMINATED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "BEFORE USE. 3 PLATES CONTAMINATED FOR THE SAME LOT NUMBER. CUSTOMER HAS DETECTED 3 PLATES CONTAMINATED FOR THE SAME LOT NUMBER."
IT WAS REPORTED THAT THE BD BBL¿ COLUMBIA AGAR PLATE WITH 5% SHEEP BLOOD WAS FOUND CONTAMINATED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "BEFORE USE. 3 PLATES CONTAMINATED FOR THE SAME LOT NUMBER. CUSTOMER HAS DETECTED 3 PLATES CONTAMINATED FOR THE SAME LOT NUMBER."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1732426 | BD BBL¿ COLUMBIA AGAR PLATE WITH 5% SHEEP BLOOD | CULTURE MEDIA, ENRICHED | KZI | BECTON DICKINSON GMBH | 1243759 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |