IFLOW
Report
- Report Number
- 3004553423-2021-01268
- Event Type
- Injury
- Date Received
- November 18, 2021
- Date of Event
- September 30, 2021
- Report Date
- October 21, 2021
- Manufacturer
- IMTMEDICAL PTE. LTD.
- Product Code
- BZO
- UDI-DI
- 07640149380767
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
RESULTS OF INVESTIGATION: THE SUSPECT DEVICE WAS NOT RETURNED FOR INVESTIGATION. VYAIRE MEDICAL WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE BUT MOST LIKELY IT IS DUE TO AN IMPROPER/ ROUGH HANDLING BASED ON SIMULATIONS AND BREAK TESTS PERFORMED.
THE SUSPECT DEVICE WAS SENT TO US FOR EVALUATION BUT WE HAVE NOT RECEIVED IT YET. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE SUSPECT DEVICE WAS SENT TO US FOR EVALUATION BUT WE HAVE NOT RECEIVED IT YET. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED BELLAVISTA - IFLOW 200S SENSOR CAME APART WHILE ON A PATIENT. FURTHERMORE, THE CUSTOMER CONFIRMED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR DUE TO PROBLEM BUT NO PATIENT HARM ASSOCIATED WITH THE EVENT.
THE CUSTOMER REPORTED BELLAVISTA - IFLOW 200S SENSOR CAME APART WHILE ON A PATIENT. FURTHERMORE, THE CUSTOMER CONFIRMED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR DUE TO PROBLEM BUT NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1737352 | IFLOW | SET,TUBING AND SUPPORT, VENTILATOR (W/HARNESS) | BZO | IMTMEDICAL PTE. LTD. | IFLOW 200 S ADULT/PEDIATRIC PROXIMAL FLOW SENSOR | 4620 | 07640149380767 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Male | Required Intervention |