FDA Adverse Event Injury Summary report: N

IFLOW

MDR report key: 12836993 · Received November 18, 2021

Report

Report Number
3004553423-2021-01268
Event Type
Injury
Date Received
November 18, 2021
Date of Event
September 30, 2021
Report Date
October 21, 2021
Manufacturer
IMTMEDICAL PTE. LTD.
Product Code
BZO
UDI-DI
07640149380767
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

RESULTS OF INVESTIGATION: THE SUSPECT DEVICE WAS NOT RETURNED FOR INVESTIGATION. VYAIRE MEDICAL WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE BUT MOST LIKELY IT IS DUE TO AN IMPROPER/ ROUGH HANDLING BASED ON SIMULATIONS AND BREAK TESTS PERFORMED.

Additional Manufacturer Narrative · 0

THE SUSPECT DEVICE WAS SENT TO US FOR EVALUATION BUT WE HAVE NOT RECEIVED IT YET. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THE SUSPECT DEVICE WAS SENT TO US FOR EVALUATION BUT WE HAVE NOT RECEIVED IT YET. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED BELLAVISTA - IFLOW 200S SENSOR CAME APART WHILE ON A PATIENT. FURTHERMORE, THE CUSTOMER CONFIRMED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR DUE TO PROBLEM BUT NO PATIENT HARM ASSOCIATED WITH THE EVENT.

Description of Event or Problem · 0

THE CUSTOMER REPORTED BELLAVISTA - IFLOW 200S SENSOR CAME APART WHILE ON A PATIENT. FURTHERMORE, THE CUSTOMER CONFIRMED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR DUE TO PROBLEM BUT NO PATIENT HARM ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1737352 IFLOW SET,TUBING AND SUPPORT, VENTILATOR (W/HARNESS) BZO IMTMEDICAL PTE. LTD. IFLOW 200 S ADULT/PEDIATRIC PROXIMAL FLOW SENSOR 4620 07640149380767

Patients

Seq Age Sex Outcome Treatment
1 45 YR Male Required Intervention