FDA Adverse Event Injury Summary report: N

TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME

MDR report key: 12836184 · Received November 18, 2021

Report

Report Number
1037905-2021-00639
Event Type
Injury
Date Received
November 18, 2021
Date of Event
September 20, 2021
Report Date
January 25, 2022
Manufacturer
COOK ENDOSCOPY
Product Code
KNS
UDI-DI
10827002225555
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER; OCCUPATION: DISTRIBUTOR. PMA/510(K) # K172665. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER: "DO NOT OVER FLEX OR BOW TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE OR CAUSE CUTTING WIRE TO BREAK." IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADDITIONAL PRESSURE IS APPLIED, THIS CAN CONTRIBUTE TO SEPARATION OF THE CATHETER AND CUTTING WIRE SECURING COMPONENT. THE INSTRUCTIONS FOR USE CAUTION THE USER: "ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING SPHINCTEROTOME." THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER: "UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME. CAREFULLY REMOVE PRECURVED STYLET FROM CANNULATING TIP." THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

INITIAL REPORTER; OCCUPATION: DISTRIBUTOR. PMA/510(K) # K172665. INVESTIGATION EVALUATION: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED IN AN UNKNOWN POUCH. THE LOT NUMBER WAS NOT PROVIDED. OUR LABORATORY EVALUATION OF THE RETURNED DEVICE SAID TO BE INVOLVED CONFIRMED THE REPORT. A VISUAL EXAMINATION OF THE DISTAL END OF THE RETURNED DEVICE SHOWED THAT THE CUTTING WIRE HAD BROKEN. A PORTION OF THE CUTTING WIRE MEASURING APPROXIMATELY 0.3CM WAS MISSING AND NOT INCLUDED IN THE RETURN OF THE DEVICE. THE CUTTING WIRE REMAINING WITHIN THE CATHETER SHOWS EVIDENCE OF CAUTERY APPLICATION AND LIQUID WAS OBSERVED IN THE CATHETER. THE CATHETER SHOWS NO KINKS/BENDS. THE CUTTING WIRE EXHIBITS SLIGHT EVIDENCE OF A CAUTERY APPLICATION (BLACKENING OF THE CUTTING WIRE WAS NOTED). A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. A BROKEN CUTTING WIRE CAN OCCUR IF THE TIP OF THE DEVICE IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER: "DO NOT OVER FLEX OR BOW TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE OR CAUSE CUTTING WIRE TO BREAK." CUTTING WIRE BREAKAGE CAN OCCUR IF THE HANDLE IS MANIPULATED WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO: ¿UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME." THE USER IS THEN INSTRUCTED TO: ¿CAREFULLY REMOVE PRECURVED STYLET WIRE FROM CANNULATING TIP." THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO THE SPHINCTEROTOME AND RENDER IT INOPERABLE." CUTTING WIRE BREAKAGE CAN ALSO OCCUR IF THE CUTTING WIRE MAKES CONTACT WITH THE ENDOSCOPE WHEN ELECTROSURGICAL CURRENT IS APPLIED TO PERFORM THE SPHINCTEROTOMY. THE INSTRUCTIONS FOR USE CAUTION THE USER: "WHEN APPLYING CURRENT, ENSURE CUTTING WIRE IS COMPLETELY OUT OF ENDOSCOPE." THE INSTRUCTIONS CAUTION THE USER: ¿CONTACT OF CUTTING WIRE WITH ENDOSCOPE MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY, A BROKEN CUTTING WIRE AND/OR DAMAGE TO ENDOSCOPE." IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADDITIONAL PRESSURE IS APPLIED, THIS COULD CONTRIBUTE TO CUTTING WIRE BREAKAGE. THE INSTRUCTIONS FOR USE CAUTION THE USER: "ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING SPHINCTEROTOME." THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. PRIOR TO DISTRIBUTION, ALL TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. IT WAS REPORTED THAT WHEN THE DOCTORS WERE GOING TO USE THE PAPILLOTOME TO MAKE THE CUT, THE WIRE CAME OFF AND DEFORMED AT THE TIP. A MILD PERFORATION WAS NOTED. IT IS UNKNOWN WHAT OR IF INTERVENTION WAS PERFORMED. COOK'S INTERPRETATION OF THE DEVICE MALFUNCTION THAT OCCURRED IS THAT THE CUTTING WIRE SECURING COMPONENT SEPARATED FROM THE CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. A MILD PERFORATION WAS NOTED. IT IS UNKNOWN WHAT OR IF INTERVENTION WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1736488 TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS COOK ENDOSCOPY W4461736 10827002225555

Patients

Seq Age Sex Outcome Treatment
1 Female Required Intervention ENDOSCOPE, UNKNOWN MAKE OR MODEL