FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12834097 · Received November 18, 2021

Report

Report Number
9610877-2021-50360
Event Type
Malfunction
Date Received
November 18, 2021
Date of Event
November 5, 2021
Report Date
November 18, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
ODG
UDI-DI
04961333239436
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG38-J10UT-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K200090. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1733642 PENTAX DIGITAL LINEAR ULTRASOUND SCOPE ODG HOYA CORPORATION PENTAX TOKYO OFFICE EG38-J10UT 04961333239436

Patients

Seq Age Sex Outcome Treatment
1 Unknown