FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12834097
·
Received November 18, 2021
Report
- Report Number
- 9610877-2021-50360
- Event Type
- Malfunction
- Date Received
- November 18, 2021
- Date of Event
- November 5, 2021
- Report Date
- November 18, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- ODG
- UDI-DI
- 04961333239436
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG38-J10UT-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K200090. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1733642 | PENTAX | DIGITAL LINEAR ULTRASOUND SCOPE | ODG | HOYA CORPORATION PENTAX TOKYO OFFICE | EG38-J10UT | 04961333239436 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |