MAKOTO INTRAVASCULAR IMAGING SYSTEM¿
Report
- Report Number
- 3015551113-2021-00001
- Event Type
- Malfunction
- Date Received
- November 17, 2021
- Date of Event
- November 11, 2021
- Report Date
- February 25, 2022
- Manufacturer
- INFRAREDX, INC.
- Product Code
- IYO
- UDI-DI
- 00857595006432
- PMA / PMN Number
- K163345
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
INVESTIGATION OF THE PRODUCT PROBLEM CONCLUDED THAT DUE TO THE DESIGN OF THE CART, A CASTER COMING OFF THE CONSOLE DID NOT PRESENT A RISK OF TIPPING/MECHANICAL HAZARD. MOREOVER, CASTER SCREWS ON CARTS BUILT SINCE 2020 ARE SUBJECT TO APPLICATION OF THREAD-LOCKING ADHESIVE AND A TORQUE SPECIFICATION. THE CART INVOLVED IN THIS COMPLAINT WAS BUILT IN 2016 PRIOR TO THE USE OF THREAD-LOCKING ADHESIVE AND A TORQUE SPECIFICATION. IN RESPONSE TO THIS COMPLAINT, THE TVC-MC10 FIELD SERVICE MANUAL WAS REVISED TO INCLUDE A CASTER BOLT CHECK AT PREVENTATIVE MAINTENANCE. IN ADDITION, THE DESIGN FMEA WAS UPDATED TO ADD AN ADDITIONAL ATTRIBUTE SPECIFICATION REGARDING CONSOLE CASTERS.
ONE OF THE FRONT WHEEL CASTERS OF THE MC10 IMAGING SYSTEM WAS REPORTED TO HAVE ALMOST FALLEN OFF OF THE CONSOLE WHILE IT WAS BEING MOVED. NO PATIENT INVOLVEMENT WAS REPORTED.