FDA Adverse Event Malfunction Summary report: N

MAKOTO INTRAVASCULAR IMAGING SYSTEM¿

MDR report key: 12832643 · Received November 17, 2021

Report

Report Number
3015551113-2021-00001
Event Type
Malfunction
Date Received
November 17, 2021
Date of Event
November 11, 2021
Report Date
February 25, 2022
Manufacturer
INFRAREDX, INC.
Product Code
IYO
UDI-DI
00857595006432
PMA / PMN Number
K163345
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION OF THE PRODUCT PROBLEM CONCLUDED THAT DUE TO THE DESIGN OF THE CART, A CASTER COMING OFF THE CONSOLE DID NOT PRESENT A RISK OF TIPPING/MECHANICAL HAZARD. MOREOVER, CASTER SCREWS ON CARTS BUILT SINCE 2020 ARE SUBJECT TO APPLICATION OF THREAD-LOCKING ADHESIVE AND A TORQUE SPECIFICATION. THE CART INVOLVED IN THIS COMPLAINT WAS BUILT IN 2016 PRIOR TO THE USE OF THREAD-LOCKING ADHESIVE AND A TORQUE SPECIFICATION. IN RESPONSE TO THIS COMPLAINT, THE TVC-MC10 FIELD SERVICE MANUAL WAS REVISED TO INCLUDE A CASTER BOLT CHECK AT PREVENTATIVE MAINTENANCE. IN ADDITION, THE DESIGN FMEA WAS UPDATED TO ADD AN ADDITIONAL ATTRIBUTE SPECIFICATION REGARDING CONSOLE CASTERS.

Description of Event or Problem · 0

ONE OF THE FRONT WHEEL CASTERS OF THE MC10 IMAGING SYSTEM WAS REPORTED TO HAVE ALMOST FALLEN OFF OF THE CONSOLE WHILE IT WAS BEING MOVED. NO PATIENT INVOLVEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1731517 MAKOTO INTRAVASCULAR IMAGING SYSTEM¿ SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC IYO INFRAREDX, INC. TVC-MC10 00857595006432

Patients

Seq Age Sex Outcome Treatment
1 Unknown