FDA Adverse Event Malfunction Summary report: N

BD PEGASUS YEL 24GAX0.75IN PRN-CAPY NON-PVC

MDR report key: 12832622 · Received November 17, 2021

Report

Report Number
3014704491-2021-00273
Event Type
Malfunction
Date Received
November 17, 2021
Date of Event
September 30, 2021
Report Date
October 29, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1152550. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING A BD PEGASUS YEL 24GAX0.75IN PRN-CAPY NON-PVC THE NEEDLE WAS DIFFICULT TO DISENGAGE. PATIENT HAD TO HAVE CATHETER REPLACED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: THE PATIENT WAS GIVEN INTRAVENOUS INDWELLING NEEDLE ON (B)(6) 2021, AFTER SUCCESSFUL PUNCTURE, THE STEEL NEEDLE OF THE INDWELLING NEEDLE COULD NOT BE REMOVED, SO IT WAS IMMEDIATELY REPLACED AND INDWELLED AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1725497 BD PEGASUS YEL 24GAX0.75IN PRN-CAPY NON-PVC INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1152550

Patients

Seq Age Sex Outcome Treatment
1 Unknown