FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 12832354 · Received November 17, 2021

Report

Report Number
3013756811-2021-125021
Event Type
Malfunction
Date Received
November 17, 2021
Date of Event
October 30, 2021
Report Date
November 17, 2021
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. A SYSTEM CHECK WAS PERFORMED BY TANDEM TECHNICAL SUPPORT AND THE OCCLUSION WAS FOUND TO BE WITHIN THE CARTRIDGE. CUSTOMER RESUMED INSULIN DELIVERY. CUSTOMER'S BLOOD GLUCOSE WAS 110-146 MG/DL

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1724620 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE

Patients

Seq Age Sex Outcome Treatment
1 46 YR Female INSULIN: NOVOLOG, INFUSION SET: AUTOSOFT XC