FDA Adverse Event Death Summary report: N

MIS TI CFX FEN POLY 7X45

MDR report key: 12830212 · Received November 17, 2021

Report

Report Number
1526439-2021-02428
Event Type
Death
Date Received
November 17, 2021
Date of Event
September 15, 2021
Report Date
September 15, 2021
Manufacturer
DEPUY SPINE INC
Product Code
PML
UDI-DI
10705034411194
PMA / PMN Number
K160879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL PROCODE: MNI. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. FAX NUMBER REPORTED AS (B)(6). WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A LUMBAR SPINAL REVISION ON (B)(6) 2021, THE PATIENT PASSED AWAY. THE PROCEDURE WAS INTENDED TO REVISE A T12-L5 CONSTRUCT AND EXTEND THE CONSTRUCT TO T9. THE REVISION WAS TO ADDRESS SCREW PULL OUT ON THE RIGHT SIDE OF T12 AND L1. THE PATIENT HAD PREVIOUS LUMBAR SURGERIES. RADIOGRAPHS INDICATED THE PATIENT MAY HAVE ANTERIOR LUMBAR INTERBODY FUSIONS (ALIFS) AND POSTERIOR FIXATION FROM L2-L5 WITH THE CLICK-X SYSTEM. A SUBSEQUENT SURGERY EXTENDED THE CONSTRUCT TWO ADDITIONAL LEVELS TO T12. IN THIS PROCEDURE, THE SURGEON DECIDED TO EXTEND THE CONSTRUCT THREE ADDITIONAL LEVELS TO T9 AND TO AUGMENT THE VERTEBRAL BODIES WITH CONFIDENCE CEMENT VIA VIPER FENESTRATED SCREWS. EACH LEVEL RECEIVED ABOUT 4-5 CC OF CEMENT (~2 CC PER SCREW). TOTAL CEMENT USED IS APPROXIMATELY 15 CC. THE PROCEDURE WENT AS PLANNED. THE WOUND CLOSURE PROCESS WAS ALMOST COMPLETE, WHEN THE PATIENT WENT INTO CARDIAC ARREST AND DIED. THIS REPORT IS FOR A MIS TI CFX FEN POLY 7X45. THIS IS REPORT 9 OF 10 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1726120 MIS TI CFX FEN POLY 7X45 BONE CEMENT, POSTERIOR SCREW AUGMENTATION PML DEPUY SPINE INC 10705034411194

Patients

Seq Age Sex Outcome Treatment
1 75 YR Female Death CONFIDENCE KIT, NO NEEDLES| CONFIDENCE KIT, NO NEEDLES| CONN O/C ANGLD 5.5-6.35X5.5TI| CONN O/O SD TOP NTCH 5.5X5.5 T| MIS TI CFX FEN POLY 6X45| MIS TI CFX FEN POLY 7X45| MIS TI CFX FEN POLY 7X45| MIS TI CFX FEN POLY 7X45| MIS TI CFX FEN POLY 7X45| SFX,5.5,TI, MED, SIZE A4| SINGLE-INNER SET SCREW| TI PRCTRD ROD S 5.5 MM| VIPER2 LORDOTIC ROD-100 MM| VIPER2 LORDOTIC ROD-150 MM