FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 12828476 · Received November 17, 2021

Report

Report Number
1221359-2021-03509
Event Type
Malfunction
Date Received
November 17, 2021
Date of Event
October 12, 2021
Report Date
February 28, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M164589 AND M163713 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: M164589, TEST BASE PART NUMBER 190-430 / LOT: M164589 AND TEST KIT PART NUMBER 190-000 / LOT: M163713, TEST BASE PART NUMBER 190-430 / LOT: M163713. THE LOTS MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M164589 AND M163713 SHOWED THAT THE COMPLAINT RATES ARE (B)(4), RESPECTIVELY. ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE; SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED (3) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED (B)(6) 2021 ON DIRECT TESTED NASAL SWAB SAMPLES AND GENERATED A NEGATIVE RESULTS WITH TWO DIFFERENT LOTS (LOT M164589 AND LOT M163713). THIS MFR. REPORT ADDRESSES PATIENT 1 OF 3. THE CUSTOMER REPORTED (1) ONE FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED (B)(6) 2021 ON A DIRECT TESTED NASAL SWAB. NO REPEAT TESTING WAS PERFORMED. INCYTE RT-PCR CONFIRMATION TESTING ON NASAL SWABS GENERATED POSITIVE RESULTS. (CT VALUES NOT PROVIDED). THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1725626 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19, QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC.

Patients

Seq Age Sex Outcome Treatment
1 Unknown